Configure Clock

Configures the properties for synchronizing the digitizer to an external clock or sending the digitizer's clock output to be used as a synchronizing clock for other devices.

1378

Inputs/Outputs

datatype_icon

session in

Handle that identifies the NI-SCOPE instrument session as previously allocated by Initialize With Options .

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Default value: No error

datatype_icon

clock sync pulse source

The line on which either the sample clock or the one-time sync pulse is sent or received.

This line should be the same for all devices to be synchronized.
RTSI 0 0 Uses the RTSI 0 bus to send or receive the sample clock or the one-time sync pulse.
RTSI 1 1 Uses the RTSI 1 bus to send or receive the sample clock or the one-time sync pulse.
RTSI 2 2 Uses the RTSI 2 bus to send or receive the sample clock or the one-time sync pulse.
RTSI 3 3 Uses the RTSI 3 bus to send or receive the sample clock or the one-time sync pulse.
RTSI 4 4 Uses the RTSI 4 bus to send or receive the sample clock or the one-time sync pulse.
RTSI 5 5 Uses the RTSI 5 bus to send or receive the sample clock or the one-time sync pulse.
RTSI 6 6 Uses the RTSI 6 bus to send or receive the sample clock or the one-time sync pulse.
PFI 1 7 Uses the PFI 1 terminal to send or receive the sample clock or the one-time sync pulse.
PFI 2 8 Uses the PFI 2 terminal to send or receive the sample clock or the one-time sync pulse.
Internal 9 Uses an internal line to send or receive the sample clock or the one-time sync pulse.

Default value: Internal

datatype_icon

master enabled

A Boolean value that specifies whether the device is a master or a slave.

The master device is typically the originator of the trigger signal and the clock sync pulse. For a standalone device, set this parameter to False.

True Sets the specified device as a master.
False Sets the specified device as a slave.
datatype_icon

input clock source

The input source for the PLL reference clock to which the digitizer is phase-locked.

None 0 No input source for the PLL reference clock.
RTSI Clock 1 Uses the RTSI clock as the input source for the PLL reference clock.
PFI 0 6 Uses the PFI 0 input as the input source for the PLL reference clock.
PFI 1 2 Uses the PFI 1 input as the input source for the PLL reference clock.
PFI 2 3 Uses the PFI 2 input as the input source for the PLL reference clock.
PXI Clock 4 Uses the clock of a PXI chassis as the input source for the PLL reference clock.
External 5 Uses an external clock as the input source for the PLL reference clock.
Clock In 7 Uses the ClkIn connector as the input source for the PLL reference clock.

Default value: None

datatype_icon

output clock source

The output source for the PLL reference clock to which another digitizer's sample clock can be phase-locked.

datatype_icon

session out

Reference to the NI-SCOPE instrument session to pass to the next node in the program.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior