Device Specific:Vector Signal Transceiver:Triggers:Synchronization:Sync Ref Trigger Master Property
- Updated2024-05-08
- 1 minute(s) read
Device Specific:Vector Signal Transceiver:Triggers:Synchronization:Sync Ref Trigger Master Property
Short Name: Sync Ref Trigger Master
Property of niRFSA
Specifies whether the device is the master for synchronizing the shared Reference Trigger between multiple devices. The master device distributes the synchronized Reference Trigger to all devices in the system through the Reference Trigger distribution line.
When synchronizing the Reference Trigger, one device must always be designated as the master. When the device is configured as a master, it actively drives the Reference Trigger distribution line. When the device is configured as a slave, set the Reference Trigger Type property to Digital Edge, and the Reference Trigger Digital Edge Source property to Sync_Ref.
Refer to the Synchronization Using NI-RFSA and NI-RFSG topic appropriate to your device in the NI RF Vector Signal Analyzers Help for more information about device synchronization for vector signal transceivers.
Default Value: FALSE
Supported Devices: PXIe-5644/5645/5646
TRUE | The device is the master device for synchronizing the Reference Trigger. |
FALSE | The device is not the master device for synchronizing the Reference Trigger. |
Related Topics
Remarks
The following table lists the characteristics of this property.
Datatype | ![]() |
Permissions | Read/Write |
High-level VIs | N/A |
Channel-based | No |
Resettable | Yes |