Configure Compression Point VI
- Updated2024-12-03
- 2 minute(s) read
Defines the compression point used for the PxdB measurement. If you are not using PxdB measurement, you do not need to use this VI.

Inputs/Outputs
PAE Reference In
—
Specifies the session reference handle of the PAE measurement session.
Compression Point (dB)
—
Specifies which compression point the PxdB algorithm will calculate. This is the difference, in dB, between the gain measured at the compression point and the gain measured within the linear range of the DUT. The default value is 1 dB for P1dB.
Maximum Linear Range (dBm)
—
Specifies the maximum DUT input power level that is in the linear range of the DUT. This value must be higher than the initial power and lower than stop power from the PRFMsc Configure RF Power Sweep VI. The default value is -5 dBm.
error in
—
Error conditions that occur before this node runs. This input provides standard error in functionality.
PAE Reference Out
—
Returns the session reference handle of the PAE measurement session.
error out
—
Error information. This output provides standard error out functionality. |