Self Calibrate
- Updated2023-02-17
- 2 minute(s) read
Self Calibrate
Self calibrates the device.
Inputs/Outputs

session in
Instrument handle that identifies your calibration session.
This parameter is obtained from the niHSDIO Initialize External Calibrationnode

error in
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Default value: No error

session out
Session handle that passes a reference to your calibration session to the next node.

error out
Error information.
The node produces this output according to standard error behavior.
Calibrating your hardware
On the PXIe-6555/6556, self-calibration may take between 15 and 20 minutes. Do not disable the device or LabVIEW during this time. You must use this node to self-calibrate your PXIe-6555/6556.
During self-calibration of PXIe-6555/6556 devices, the device conducts an analog voltage calibration and a digital timing calibration.
During self-calibration of the PXI-6541/6542,PXI-6551/6552, and PXI-6561/6562 devices, the voltage-controlled crystal oscillator (VCXO) phase D/A converters are recalibrated.