Reset Device
- Updated2023-02-17
- 2 minute(s) read
Reset Device
Resets the device to its initial state and reloads the FPGA. Use this node to re-enable your device if it has disabled itself because the device temperature has risen above its optimal operating temperature.
All channels and front panel terminals are put into a high-impedance state. All software attributes are reset to their initial values. The entire contents of the FPGA and EEPROM files are reloaded.
During a device reset, signal routes between this and other devices are released, regardless of which device created the route. For instance, a trigger signal being exported to a PXI trigger line and used by another device is no longer exported.
Generally, using the Resetnode is acceptable in place of the Reset Devicenode. The Resetnode executes more quickly.
Inputs/Outputs

session in
Instrument session obtained from the Initialize Acquisition Sessionnode, the Initialize Generation Sessionnode, or the Initialize External Calibrationnode.

error in
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Default value: No error

session out
Handle that identifies the session in all subsequent node calls.
The session out parameter is obtained from the Initialize Acquisition Sessionnode or the Initialize Generation Sessionnode.

error out
Error information.
The node produces this output according to standard error behavior.