Configure Calibration Device State

Configures the internal routing of the device so that you can perform specific calibration steps.

Inputs/Outputs

datatype_icon

session in

Instrument session obtained from the Initialize Acquisition Sessionnode, the Initialize Generation Sessionnode, or the Initialize External Calibrationnode.

datatype_icon

calibration reference

Parameter that specifies which internal routing path to configure based on the calibration type.

Refer to the calibration procedure document for your device for the correct value to use.

datatype_icon

calibration type

The section of the external calibration procedure you are trying to execute.

Refer to the calibration procedure document for your device for the correct value to use.

datatype_icon

connector

Parameter that selects whether to access the PMU pins on the AUX I/O connector or the REMOTE SENSE connector.

AUX I/O

Accesses the PMU pins on the AUX I/O connector. This is the default value.

Remote Sense

Accesses the PMU pins on the REMOTE SENSE connector.

Using this parameter with the PXIe-6555/6556

This parameter is valid only on the PXIe-6556, which has both the AUX I/O and REMOTE SENSE connectors. If you are operating an PXIe-6555, you must select Remote Sense because the PXIe-6555 does not have an AUX I/O connector. If you select AUX I/O while using an PXIe-6555, NI-HSDIO returns an error.

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Default value: No error

datatype_icon

session out

Session handle that passes a reference to your calibration session to the next node.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior