Close External Calibration
- Updated2023-02-17
- 2 minute(s) read
Close External Calibration
Closes an NI-HSDIO external calibration session and, if specified, stores the new calibration constants and calibration data in the onboard EEPROM.
Note Whether you commit or cancel, the device is reset and the FPGA is reloaded afterwards.
Inputs/Outputs

session in
Instrument handle that identifies your calibration session.
This parameter is obtained from the niHSDIO Initialize External Calibrationnode

calibration action
The action to perform at closing.
| Commit | 62 | The new calibration constants and data determined during the external calibration session are stored in the onboard EEPROM, given that the calibration was complete and passed successfully. |
| Cancel | 63 | No changes are made to the calibration constants and data in the EEPROM. |
Default value: Commit

error in
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Default value: No error

error out
Error information.
The node produces this output according to standard error behavior.