niFgen Cal Adjust Oscillator Frequency VI
- Updated2025-10-07
- 2 minute(s) read
Calculates calibration constants pertaining to the oscillator. During external calibration, you can generate sine waves and take measurements of the resulting output frequencies. The desired and measured frequencies are passed to this VI so that NI-FGEN can calculate the appropriate calibration constants and, when the calibration session is committed, store them in the onboard EEPROM.

Inputs/Outputs
Instrument Handle
—
Instrument Handle identifies your instrument session. Instrument Handle is obtained from the niFgen Init Ext Cal VI.
Desired Frequency In Hz
—
Desired Frequency In Hz specifies the expected frequency of the output waveform.
Measured Frequency In Hz
—
Measured Frequency In Hz specifies the actual (measured) frequency of the output waveform.
error in (no error)
—
error in describes error conditions that occur before this node runs. This input provides standard error in functionality. Default value: no error
Instrument Handle Out
—
Instrument Handle Out passes a reference to your instrument session to the next VI.
error out
—
error out contains error information. This output provides standard error out functionality. |
Instrument Handle
—
Desired Frequency In Hz
—
error in (no error)
—
Instrument Handle Out
—
error out
—