niFgen Cal Adjust Direct Path Output Impedance VI
- Updated2025-10-07
- 3 minute(s) read
Calculates calibration constants pertaining to direct analog path output impedance. During external calibration, you can put the device in different configurations and take measurements of the resulting output voltage across different loads. The configuration data, as well as the measurements, are passed to this VI so that NI-FGEN can calculate the appropriate calibration constants and, when the calibration session is committed, store them in the onboard EEPROM.

Inputs/Outputs
Load Impedance
—
Load Impedance specifies the impedance of the load across which the measurement passed in as Measured Voltage Across Load is taken.
Configuration
—
Configuration specifies the configuration of the device for this stage of calibration.
Instrument Handle
—
Instrument Handle identifies your instrument session. Instrument Handle is obtained from the niFgen Init Ext Cal VI.
Channel Name
—
Channel Name specifies the channel that this VI uses.
Measured Source Voltage
—
Measured Source Voltage specifies the analog output voltage measured across a very high-impedance load.
error in (no error)
—
error in describes error conditions that occur before this node runs. This input provides standard error in functionality. Default value: no error
Measured Voltage Across Load
—
Measured Voltage Across Load specifies the analog output voltage measured across the load impedance specified in Load Impedance.
Instrument Handle Out
—
Instrument Handle Out passes a reference to your instrument session to the next VI.
error out
—
error out contains error information. This output provides standard error out functionality. |
Load Impedance
—
Configuration
—
Instrument Handle
—
Channel Name
—
error in (no error)
—
Instrument Handle Out
—
error out
—