niFgen Cal Adjust Cal ADC VI
- Updated2025-10-07
- 3 minute(s) read
Calculates calibration constants pertaining to the gain and offset of the onboard calibration ADC. During external calibration, you can generate voltages and measure them both externally and with the calibration ADC. The measured voltages are passed to this VI so that NI-FGEN can calculate the appropriate calibration constants and, when the calibration session is committed, store them in the onboard EEPROM.

Inputs/Outputs
Channel Name
—
Channel Name specifies the channel that this VI uses.
Configuration
—
Configuration specifies the configuration of the device for this stage of calibration.
Instrument Handle
—
Instrument Handle identifies your instrument session. Instrument Handle is obtained from the niFgen Init Ext Cal VI.
Voltages Measured Externally
—
Voltages Measured Externally is an array of analog output voltages measured with an external instrument.
Voltages Measured with Cal ADC
—
Voltages Measured with Cal ADC is an array of analog output voltages measured with the onboard calibration ADC.
error in (no error)
—
error in describes error conditions that occur before this node runs. This input provides standard error in functionality. Default value: no error
Instrument Handle Out
—
Instrument Handle Out passes a reference to your instrument session to the next VI.
error out
—
error out contains error information. This output provides standard error out functionality. |
Channel Name
—
Configuration
—
Instrument Handle
—
Voltages Measured Externally
—
error in (no error)
—
Instrument Handle Out
—
error out
—