niDMM_SelfCal
- Updated2023-02-21
- 1 minute(s) read
ViStatus = niDMM_SelfCal(ViSession Instrument_Handle)
Purpose
For the NI 4080/4081/4082 and the NI 4070/4071/4072, executes the self-calibration routine to maintain measurement accuracy.
![]() |
Note This function calls niDMM_reset, and any configurations previous to the call will be lost. All attributes will be set to their default values after the call returns. |
Parameters
Input | ||
Name | Type | Description |
Instrument_Handle | ViSession | Identifies a particular instrument session. You obtain the Instrument_Handle parameter from niDMM_init or niDMM_InitWithOptions. The default is None. |
Return Value
Name | Type | Description |
Status | ViStatus | Reports the Status of this operation. To obtain a text description of the status code, call niDMM_GetErrorMessage. To obtain additional information concerning the error condition, use niDMM_GetError. |