niDCPower Initialize External Calibration VI
- Updated2025-04-18
- 2 minute(s) read
If password is valid, this VI creates a new IVI instrument driver session to the device specified in resource name and returns an instrument handle you use to identify the device in all subsequent NI-DCPower VI calls. This VI also sends initialization commands to set the device to the state necessary for the operation of NI-DCPower.
Opening a calibration session always performs a reset. Refer to the calibration procedure for the device you are calibrating for detailed instructions on the appropriate use of this VI.

Inputs/Outputs
![]() resource name specifies the resource name assigned by Measurement & Automation Explorer (MAX), for example, PXI1Slot3, where PXI1Slot3 is an instrument resource name. resource name can also accept a logical IVI name for a device. ![]() password specifies the password for opening a calibration session. password can be a maximum of four alphanumeric characters. The initial password is factory configured to NI. ![]() error in describes error conditions that occur before this node runs. This input provides standard error in functionality. ![]() instrument handle out passes the handle used to identify the session in all subsequent NI-DCPower VI calls. ![]() error out contains error information. This output provides standard error out functionality. |