ConductionVoltageMode
- Updated2025-10-06
- 1 minute(s) read
Specifies whether the conduction voltage feature is enabled on the specified channel(s).
Syntax
Namespace: NationalInstruments.ModularInstruments.NIDCPower
public DCPowerConductionVoltageMode ConductionVoltageMode { get; set; }
Remarks
Whether the conduction voltage feature is enabled on the specified channel(s).
When the conduction voltage feature is enabled:
The instrument will not begin sinking on the specified channel(s) until the voltage at the input of the specified channel(s) rises above ConductionVoltageOnThreshold.
The instrument will stop sinking on the specified channel(s) if the voltage at the input of the specified channel(s) falls below ConductionVoltageOffThreshold.
When the conduction voltage feature is disabled, the instrument will start sinking on the specified channel(s) regardless of the voltage at the input of the specified channel(s).
This property is not supported by all instruments. Refer to the Supported Properties by Device topic in the NI DC Power Supplies and SMUs Help for information about supported instruments.