Balances between settling time and maximum measurement time by specifying the maximum time delay between when a range change occurs and when measurements resume.

Syntax

Namespace: NationalInstruments.ModularInstruments.NIDCPower

public double AutorangeMaximumDelayAfterRangeChange { get; set; }

Remarks

The default value is 0.5 seconds.

Valid Values: The minimum and maximum values of this property are hardware-dependent.

  • PXIe-4135/4136/4137: 0 to 9 seconds

  • PXIe-4138/4139: 0 to 9 seconds

  • PXIe-4147: 0 to 9 seconds

  • PXIe-4162/4163: 0 to 0.1 seconds

Note

This property is not supported by all instruments. Refer to the Supported Properties by Device topic in the NI DC Power Supplies and SMUs Help for information about supported instruments.