NI-9425 Specifications
- Updated2024-11-25
- 4 minute(s) read
NI-9425 Specifications
Introduction
In this document, the NI-9425 with spring terminal and NI-9425 with DSUB are referred to inclusively as the NI-9425. The information in this document applies to all versions of the NI-9425 unless otherwise specified.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted.
NI-9425 Pinout
| Signal | Description |
|---|---|
| COM | Common reference connection to isolated ground |
| DI | Digital input signal connection |
| NC | No connection |
Input Characteristics
Number of channels | 32 digital input channels | ||||||||||||||||||||||||||
Input type | Sinking | ||||||||||||||||||||||||||
| |||||||||||||||||||||||||||
Input impedance | 30 kΩ ± 5% | ||||||||||||||||||||||||||
| |||||||||||||||||||||||||||
Hold time[1]1 Hold time is the amount of time input signals must be stable after initiating a read from the module. | 0 μs minimum | ||||||||||||||||||||||||||
Setup time[2]2 Setup time is the amount of time input signals must be stable before reading from the module. | 1 μs minimum | ||||||||||||||||||||||||||
| |||||||||||||||||||||||||||
MTBF | 1,256,699 hours at 25 °C; Bellcore Issue 2, Method 1, Case 3, Limited Part Stress Method | ||||||||||||||||||||||||||
Power Requirements
| |||||||
| |||||||
Physical Characteristics
| |||||||||||||
| |||||||||||||
| |||||||||||||
NI-9425 with Spring Terminal Safety Voltages
Connect only voltages that are within the following limits:
Channel-to-COM | 60 V DC | ||||||||||||
| |||||||||||||
NI-9425 with DSUB Safety Voltages
Connect only voltages that are within the following limits:
Channel-to-COM | 60 V DC | ||||||||||||||
| |||||||||||||||
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 2,000 m | ||||||||||||
| |||||||||||||
To meet these shock and vibration specifications, you must panel mount the system.
Calibration
Calibration interval | 1 year |
1 Hold time is the amount of time input signals must be stable after initiating a read from the module.
2 Setup time is the amount of time input signals must be stable before reading from the module.
3 The update/transfer time is valid when the module is used in a CompactRIO system. When used in other systems, driver software and system latencies impact this time.