Configure Trigger Delay

Configures the delay between the trigger and the actual measurement.

You can configure a trigger delay only when you set the trigger source input of Configure Trigger Source to Internal or External.

Inputs/Outputs

datatype_icon

session in

Instrument session obtained from Initialize With Options.

datatype_icon

error in

Error conditions that occur before this node runs.

The node responds to this input according to standard error behavior.

Standard Error Behavior

Default value: No error

datatype_icon

trigger delay enabled

Boolean value that specifies if the RF power meter uses a trigger delay when making a triggered measurement.

True The RF power meter uses a delay.
False The RF power meter does not use a delay.

Default value: False

datatype_icon

trigger delay time

Trigger delay time in seconds.

A negative delay means the sensor takes measurements before the trigger occurs.

Valid Range: -5 x 10-3 to 10.0

Default value: 0.0

datatype_icon

session out

Reference to your instrument session passed to the next node.

datatype_icon

error out

Error information.

The node produces this output according to standard error behavior.

Standard Error Behavior

Hardware Support

This node supports the following hardware:

  • USB-5681
  • USB-5683
  • USB-5684