Convert RTD Reading
- Updated2023-02-17
- 2 minute(s) read
Convert RTD Reading
Converts an RTD voltage into a temperature value in degrees Celsius.


Ro
The RTD resistance at 0 degrees Celsius.
Default: 100 Ω

voltage
The voltage read from the RTD.
This input supports waveform and double-precision, floating-point data types.

Iex
The excitation current used with the RTD.
Default: 0.15 µA

A
A coefficient of the Callendar Van-Dusen equation that fits your RTD.
Default: 0.00390802 — Represents the coefficient for the European curve (also called the DIN 43760 standard).

B
A coefficient of the Callendar Van-Dusen equation that fits your RTD.
Default: -5.80195E-07 — Represents the coefficient for the European curve (also called the DIN 43760 standard).

temperature
The return temperature value in degrees Celsius.