Calibrating TB-9212 Using Calibration Executive
- Updated2025-11-03
- 4 minute(s) read
Calibrating TB-9212 Using Calibration Executive
Calibrate the TB-9212 with mini-TC or the TB-9212 with screw terminal using Calibration Executive.
For more information on these terminal block variants, refer to the TB-9212 Calibration Procedure on ni.com/docs.
Calibration Executive Procedure Features
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
18 minutes
Test Equipment
The following table lists the test instruments required for calibrating the TB-9212 terminal blocks.
| Instrument | Recommended Model | Requirements | |
|---|---|---|---|
| Thermocouple Calibrator | Ectron 1140A | If this instrument is unavailable, use a high precision thermocouple source with an accuracy of ≤0.06 °C and an output impedance of ≤10 Ω. | |
| C Series Module | NI-9212 | An NI-9212 module that is verified to be within product specifications. | |
| Chassis | cDAQ-9178 | — | |
| Barrier Strip Assembly | Thermocouple Wire | TT-E-24-SLE-100 | E-type SLE 24 AWG, 20 m to 30 m. |
| Mini Thermocouple Connector | Omega SMPW-E-M | E-type, must fit the output terminal of the thermocouple calibrator. | |
| Thermocouple Barrier Strips | Omega BS16A | 8 channels. | |
| Thermocouple Barrier Strip Jackets (x8) | Omega BSJ-E | E-type, 8 channels. | |
| Thermocouple Spade Lugs (x32) | Omega SLCH-20 | 16 lugs of chromel to match the E-type wires. | |
| Omega SLCO-20 | 16 lugs of constantan to match the E-type wires. | ||
| Spade-Lug Crimping Tool | Omega CRIMPING TOOL-P | Must effectively crimp the spade lugs with the 24 AWG wire. | |
| Mini Thermocouple Connector (x8) | Omega SMPW-U-M | U type Note This instrument is
required only for the TB-9212 with mini-TC. Refer to the Barrier
Strip Assembly figure. |
|
Test Conditions
The following setup and environmental conditions are required to ensure the TB-9212 meets published specifications.
- Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Verify that all connections to the device are secure.
- Use thermocouple wire for all cable connections to the TB-9212.
- Maintain an ambient temperature of 23 °C ±5 °C. The device temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at least 15 minutes to ensure that the NI-9212 measurement circuitry is at a stable operating temperature.
Device Setup
- Make sure the cDAQ-9178 power source is not connected.
- Connect the cDAQ-9178 to the system safety ground.
- Attach a ring lug to a 14 AWG (1.6 mm) wire.
- Connect the ring lug to the ground terminal on the side of the cDAQ-9178 using the ground screw.
- Attach the other end of the wire to the system safety ground.
- Install the NI-9212 in slot 8 of the cDAQ-9178 chassis. Leave slots 1 through 7 of the cDAQ-9178
chassis empty. Note When multiple NI-9212 measurement devices are configured, you may have the option to calibrate multiple TB-9212 terminal blocks in a single session. Because only one measurement device can configured as an instrument for the session, each terminal block must be connected to the selected measurement device during its calibration session.
- Connect the cDAQ-9178 chassis to your host computer.
- Connect the power source to the cDAQ-9178 chassis.
- Launch Measurements & Automation Explorer (MAX).
- Right-click the device name and select Self-Test to ensure that the device is working properly.
Barrier Strip Assembly
Test Limit Equations
The following test limits are derived from the published specifications.
| Equation Type | Equation |
|---|---|
| Temperature Accuracy |
ErrorLimits = abs(Tempchannel - TempCJC) Temperror(TTC, TCJ) = ErrorLimits Temperror(TTC, TCJ) = TE * [VMEASURED(TTC, TCJ) + VError_MODULE(TTC) + VE(TCJ + TCJ_err)] - TTC
|