Calibrate the TB-4302C terminal block using Calibration Executive.

Calibration Executive Procedure Features

Verify Only Verify & Adjust Manual Mode Optional Adjust

Approximate Test Time

7 minutes

Test Equipment

The following table lists the test instruments required for calibrating the TB-4302C.

Table 139. Test equipment for calibrating the TB-4302C
Instrument Recommended Model Requirements
DMM PXI-4071 Use a DMM that has an accuracy of 136 ppm or better when measuring 5 Ω in 4-wire mode.
4-wire Kelvin probe Keithley 5808 Use a probe capable of measuring Kelvin resistance.

Test Conditions

The following setup and environmental conditions are required to ensure the TB-4302C meets published specifications.

  • Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements
  • Maintain an ambient temperature of 23 °C ±5 °C. The device temperature will be greater than the ambient temperature.
  • Keep the relative humidity below 80%.

Device Setup

  1. Loosen the two jackscrews and remove the cover from the TB-4302C.
  2. Launch the Calibration Executive procedure. Calibration Executive guides you through the required connections.

Test Limit Equations

The following test limits are derived from the published specifications.

Table 140. TB-4302C Test Limit Equations
Equation Type Equation
Shunt Value Accuracy Verification TestLimit = TestValue ± [TestValue * (GainError / 100)]