Calibrating PXIe-6614 Using Calibration Executive
- Updated2025-11-03
- 3 minute(s) read
Calibrate the PXIe-6614 using Calibration Executive.
Calibration Executive Procedure Features
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
| Verify Only | Verify & Adjust |
|---|---|
| 3 minutes | 9 minutes |
Test Equipment
The following table lists the test equipment required to calibrate the PXIe-6614.
| Instrument | Recommended Model | Requirements |
|---|---|---|
| 10 MHz rubidium frequency standard clock or better* | Symmetricom 8040C or Datum 8040 | ±0.75 ppb† accuracy or better |
| Shielded 68-conductor cable | Sh68-68-D1 | Terminated with 68-pin female 0.050 series D-type connector |
| I/O connector block | BNC-2121, SCB-68A, or any 68-pin breakout box | — |
| BNC male-to-BNC male cable | — | 50 Ω |
| 75 Ω BNC male to 50 Ω BNC female adapter | — | 50 to 75 Ω |
- *Allow adequate time for the clock to warm up prior to calibration.
- †1 ppb = 1/1000,000,000
Test Conditions
- Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Verify that all connections to the device, including front panel connections, are secure.
- Use shielded coaxial cables or shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
- Maintain an ambient temperature of 23 °C ± 5°C.
- Keep relative humidity between 5% and 95%, non-condensing.
- Allow a warm up time of at least two hours to ensure that the measurement circuitry is at a stable operating temperature.
Device Setup
- Power off the PXI Express chassis, and install the PXIe-6614 in an available slot.
- Power on the chassis and launch MAX.
- Configure the device identifier and select Self-Test to ensure that the device is working properly.
- Connect the PXIe-6614 to the I/O connect or block using the shielded 68-conductor cable.
- Connect the clock source to PFI 39 on the PXIe-6614.
Test Limit Equations
The following test limits are derived from the published specifications.
| Equation Type | Equation |
|---|---|
| Clock Accuracy Verification and Adjustment | Frequency = 100 MHz Temperature and 1-year drift = 75 ppb Retrace error = 10 ppb Calibration measurement accuracy = 4.25 ppb As-found Limit = Frequency * (1 ± (Temperature and 1-year drift + Retrace error)) = 100 MHz * (1 ± (75 ppb + 10 ppb)) = ± 8.5 Hz As-left Limit = Frequency * (1 ± Calibration measurement accuracy) = 100 MHz * (1 ± 4.25 ppb) = ± 0.425 Hz |