Calibrating NI R Series Using Calibration Executive
- Updated2025-11-03
- 3 minute(s) read
Calibrating NI R Series Using Calibration Executive
Calibrate the NI 7841R/7842R/7845R/7846R/7847R/7851R/7852R/7853R/7854R/7855R/7856R/7857R/7858R/7867R/7868R and the PXIe-7861/7862 using Calibration Executive.
Calibration Executive Procedure Features
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Times
| Device | Verify Only | Verify & Adjust | Warmup |
|---|---|---|---|
| NI 784x/785x | 5 minutes | 10 minutes | 15 minutes |
| PXIe-7861/7862 | 10 minutes | 15 minutes | 15 minutes |
| NI 7867R/7868R | 10 minutes | 20 minutes | 15 minutes |
Test Equipment
| Required Instrument | Recommended Model | Requirements |
|---|---|---|
| Calibrator | Fluke 5700A/5730A | If this instrument is unavailable, use a calibrator that is accurate to 10 ppm. |
| DMM | NI 4070, PXIe-4080 | If this instrument is unavailable, use a multiranging 6 1/2-digit DMM with an accuracy of 40 ppm. |
| Counter | Agilent 53132A, Agilent 53220A | If this instrument is unavailable, use a counter that is accurate to 0.01%. |
| Chassis | PXI-1042, PXI-1042Q | Use with PXI modules. |
| PXIe-1062Q | Use with PXI Express modules. | |
| Low thermal Copper EMF Plug-in Cable | Fluke 5440A-7002 | Do not use standard banana cables. |
| DAQ Accessory | NI SCB-68 | Shielded I/O connector block with 68 screw terminals for easy signal connection to 68- or 100-pin R Series devices. |
| NI CB-68LP, NI CB-68LPR, NI TBX-68 | Low-cost termination accessories with 68 screw terminals for easy connection of field I/O signals to 68-pin R Series devices. | |
| Shielded MIO Cable | NI SHC68-68-RMIO | High-performance shielded cable for R Series devices. |
Test Conditions
The following setup and environmental conditions are required to ensure the NI R Series
device meets published specifications:
- Keep connections to the R Series device short. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
- Maintain an ambient temperature between 18 and 28°C. The device temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at least 15 minutes to ensure that the measurement circuitry of the device is at a stable operating temperature.
- (PXI/PXI Express) Ensure that the PXI/PXI Express chassis fan speed is set to HIGH and that the empty slots contain filler panels. For more information about maintaining adequate air circulation in your PXI/PXI Express chassis, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/docs.
Device Setup
The device must be configured in Measurement & Automation Explorer (MAX) to communicate with the NI R Series Multifunction RIO driver.
Complete the following steps to
configure the device in MAX:
- Install the NI R Series Multifunction RIO driver software. driver software.
- Power off the computer or chassis that will hold the device and install the device in an available slot.
- Power on the computer or chassis and launch MAX.
- In the left pane of MAX, expand Devices and Interfaces»NI-RIO
Devices to locate your device. Note the Resource Name for this device (for
example, R100). Note When a RIO device is configured with MAX, it is assigned a Resource Name. This Resource Name is used to communicate with the device during calibration.
- Launch the Calibration Executive procedure and complete the setup wizard.
Test Limit Equations
The following test limits are derived from the published specifications:
| Equation Type | Equation |
|---|---|
| Analog Input | TestLimits = TestValue ± [TestValue * GainError (% of Reading) + Offset (µV) + Noise and Quantization (µV)] |
| Analog Output | TestLimits = TestValue ± [TestValue * GainError (% of Reading) + Offset (µV)] |
| Counter | TestLimits = TestValue ± [TestValue * Accuracy (ppm)] |