Calibrating Isolated M/S Series Using Calibration Executive
- Updated2025-11-03
- 3 minute(s) read
Calibrating Isolated M/S Series Using Calibration Executive
Calibrate the PXI/PCI-6230, PXI/PCI-6232, PXI/PCI-6233, PXI/PCI-6236, PXI/PCI-6238, PXI/PCI-6239 and PCI-6154 using Calibration Executive.
Calibration Executive Procedure Features
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
10 minutes
Test Equipment
The following table lists the test equipment required for calibrating M/S Series Isolated devices.
| Instrument | Recommended Model | Requirements |
|---|---|---|
| Calibrator | Fluke 5700A | If this instrument is unavailable, use a voltage source that is at least 10 ppm
accurate for 16-bit voltage input devices. For 16-bit current input devices, use a current source that is at least 100 ppm accurate. |
| DMM | NI 4070 | If this instrument is unavailable, use a multiranging 6 1/2-digit DMM with an accuracy of 40 ppm. |
| Counter | Agilent 53131A | If this instrument is unavailable, use a counter with an accuracy of 0.01%. |
| Resistor | — | (NI 6232/6233/6238/6239 Devices) For counter verification, use any 1 kΩ resistor. |
| Power Supply | NI 4110 | (NI 6238/6239 Devices) For analog output verification, use any 10–30 VDC, 100
mA power supply. (NI 6232/6233/6238/6239 Devices) For counter verification, use any 5 VDC, 200 mA power supply. |
| Chassis | PXI-1042, PXI-1042Q | Use with PXI modules. |
| Low thermal Copper EMF Plug-in Cable | Fluke 5440A-7002 | Do not use standard banana cables. |
| Shielded DAQ Cable | NI DB37M-DB37F-EP, NI SH37F-37M-1 | Use with NI 6154, NI 6230, and NI 6236 devices. |
| NI SH37F-37M-1 | Use with NI 6232, NI 6233, NI 6238 and NI 6239 devices. | |
| DAQ Accessory | NI CB-37F-LP | Low profile terminal block with 37 screw terminals. |
| NI CB-37FH | Horizontal DIN-mountable terminal block with 37–pin screw terminals. | |
| NI CB-37FV | Vertical DIN-mountable terminal block with 37–pin screw terminals. | |
| NI CB-37F-HVD | 37-pin DIN rail screw terminal block, UL Recognized derated to 30 Vrms, 42.4 Vpk, or 60 VDC. |
Test Conditions
The following setup and environmental conditions are required to ensure the M/S Series Isolated device meets published specifications.
- Keep connections to the device short. Long cables and wires can act as antennas, which can pick up extra noise that might affect measurements.
- Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
- Maintain a temperature between 18 and 28 °C. The device temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at least 15 minutes to ensure that the measurement circuitry of the device is at a stable operating temperature.
- (PXI) Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters are clean, and that the empty slots contain filler panels.
Device Setup
Configure the isolated M/S Series device in Measurement & Automation Explorer (MAX) to communicate with NI-DAQmx.
- Power off the computer or chassis that will hold the device, and install the device in an available slot.
- Power on the computer and launch MAX.
- Configure the device identifier and select Self-Test to ensure that the
device is working properly. Note When a device is configured with MAX, it is assigned a device identifier. Each function call uses this identifier to determine which DAQ device to calibrate.
- Launch the Calibration Executive procedure and complete the setup wizard.