Calibrating NI-9269 Using Calibration Executive
- Updated2025-11-03
- 2 minute(s) read
Calibrate the NI-9269 using Calibration Executive.
Calibration Executive Procedure Features
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
3 minutes
Test Equipment
The following table lists the test equipment required for calibrating NI-9269 modules.
| Instrument | Recommended Model | Requirements |
|---|---|---|
| DMM | NI 4070 DMM | If this instrument is unavailable, use a multiranging 6 1/2-digit DMM with a voltage accuracy of ±400 ppm. |
| Chassis | cDAQ-9178 | — |
Test Conditions
The following setup and environmental conditions are required to ensure the NI-9269 meets published specifications.
- Keep connections to the NI-9269 as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Verify that all connections to the NI-9269 are secure.
- Use shielded copper wire for all cable connections to the NI-9269. Use twisted-pair wire to eliminate noise and thermal offsets.
- Maintain an ambient temperature of 23 °C ± 5 °C. The NI-9269 temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at least 10 minutes to ensure that the NI-9269 measurement circuitry is at a stable operating temperature.
Device Setup
- Install the NI-9269 in slot 8 of the cDAQ-9178 chassis. Leave slots 1 through 7 on the cDAQ-9178 chassis empty.
- Connect the cDAQ-9178 chassis to your host computer.
- Connect the power source to the cDAQ-9178 chassis.
- Configure the hardware using Measurement & Automation Explorer (MAX).
- Launch the Calibration Executive procedure and complete the setup wizard.
Test Limit Equations
The following test limits are derived from the published specifications found in the NI-9269 Calibration Procedure.
| Equation Type | Equation |
|---|---|
| Voltage Accuracy | VoltageAccuracyLimit = TestValue ± [(TestValue * 0.06%) + (Range * 0.015%)] |