Calibrating NI-9238 Using Calibration Executive
- Updated2025-11-03
- 2 minute(s) read
Calibrate the NI-9238 using Calibration Executive.
Calibration Executive Procedure Features:
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
| Warm-up | Verify & Adjust |
|---|---|
| 10 minutes | 6 minutes |
Test Equipment
The following table lists the test equipment required for calibrating NI-9238 modules.
| Instrument | Recommended Model | Requirements |
|---|---|---|
| Calibrator | Fluke 5500A | If this instrument is unavailable, use a high-precision voltage source with an accuracy of ≤60 ppm and offset error ≤6 µV. |
| Chassis | cDAQ-9178 | — |
Test Conditions
The following setup and environmental conditions are required to ensure the NI-9238 meets published specifications.
- Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
- Maintain an ambient temperature of 23 °C ± 5 °C. The device temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at least 10 minutes to ensure that the NI-9238 measurement circuitry is at a stable operating temperature.
Device Setup
- Install the NI-9238 in slot 8 of the cDAQ-9178 chassis. Leave slots 1 through 7 on the cDAQ-9178 chassis empty.
- Configure the hardware using Measurement & Automation Explorer (MAX).
- Launch the Calibration Executive procedure and complete the setup wizard.
Note If the NI-9238 module fails after calibration, return it to NI for repair or replacement.
Test Limit Equations
The following test limits are derived from the published specifications found in the NI-9238 Calibration Procedure.
| Equation Types | Equation |
|---|---|
| Accuracy Verification | AccuracyVerificationLimit = TestValue ± (TestValue * GainError + Range * Offset Error) |