Calibrate the FD-11601 Voltage Input Device for FieldDAQ using Calibration Executive.

Calibration Executive Procedure Features

Verify Only Adjust Only Verify & Adjust Manual Mode Optional Adjust

Approximate Test Time

Module Warmup Verify Only Verify & Adjust Adjust Only
FD-11601 10 minutes 5 minutes 10 minutes 5 minutes

Test Equipment

Table 101. Test Equipment for Calibrating the FD-11601
Instrument Recommended Model Requirements
Calibrator Fluke 5522A or 5560A High-precision voltage source with gain uncertainty of ≤50 ppm and offset uncertainty of ≤50 μV
SHM125M-Pigtail I/O Cable (0.3 m, 1 m, 3 m lengths) (x4) Part number 785950-0R3/01/03
Banana plug (optional) (x2) Use banana plugs with an opening large enough to accommodate four 22 AWG signal wires
Power supply

IP67 Power Supply, part number 147464-01

or

PS-16 Power Supply, part number 781094-01

M125F-Pigtail Power Cable (1 m, 2 m, 3 m lengths) Part number 786172-01/02/03
M125F-M125M Power Cable (0.3 m, 1 m, 3 m lengths) (optional) Part number 786173-0R3/01/03 Required for daisy-chaining multiple FieldDAQ devices
SHM128M-RJ45 Ethernet Cable (1 m, 2 m, 3 m, 10 m lengths) Part number 785944-01/02/03/10
SHM128M-SHM128M Ethernet Cable (0.3 m, 1 m, 3 m lengths) (optional) Part number 785946-0R3/01/03 Required for daisy-chaining multiple FieldDAQ devices

Test Conditions

The following setup and environmental conditions are required to ensure the FD-11601 meets published specifications.

  • Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
  • Verify that all connections to the device are secure.
  • Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to eliminate noise and thermal offsets.
  • Maintain an ambient temperature of 23 °C ±5 °C. The device temperature will be greater than the ambient temperature.
  • Keep relative humidity below 80%.
  • Allow a warm-up time of at least 10 minutes to ensure that the FieldDAQ device measurement circuitry is at a stable operating temperature.

Device Setup

To set up the device for calibration, complete the following steps:

  1. Set up the FD-11601 as outlined in the FD-11601 User Manual.
    Note NI recommends that you use a dedicated Ethernet network adapter to connect the FD-11601 to a computer or controller. Use a second Ethernet network adapter for additional network access.
  2. Configure the hardware using Measurement & Automation Explorer (MAX).
  3. Launch the Calibration Executive procedure and complete the setup wizard.
Note If the device fails after calibration, return it to NI for repair or replacement.

Test Limit Equations

The following test limits are derived from the published specifications:

TestLimits = TestValue ± ( | TestValue | × ppm of reading 1000000 + Range × ppm of range 1000000 )