Calibrating CompactDAQ Chassis Using Calibration Executive
- Updated2025-11-03
- 3 minute(s) read
Calibrating CompactDAQ Chassis Using Calibration Executive
Calibrate a cDAQ-9170, cDAQ-9171, cDAQ-9173, cDAQ-9174, cDAQ-9177, cDAQ-9178, cDAQ-9179, cDAQ-9181, cDAQ-9183, cDAQ-9184, cDAQ-9185, cDAQ-9187, cDAQ-9188, cDAQ-9189, or cDAQ-9191 chassis using Calibration Executive.
Calibration Executive Procedure Features
| Verify Only |
Verify & Adjust |
Manual Mode |
Optional Adjust |
Approximate Test Time
1 minute
Test Equipment
The following table lists the test equipment required for calibrating CompactDAQ chassis devices.
| Instrument | Recommended Model | Requirements |
|---|---|---|
| Counter | Keysight 53131A, Keysight 53220A | If this instrument is unavailable, use a counter accurate to ±5 ppm. |
| C Series Digital I/O Module | NI 9401, NI 9402 | — |
| C Series Connection Accessory | NI 9924 | (NI 9401) 25-pin DSUB to screw-terminal connector block. |
Test Conditions
The following setup and environmental conditions are required to ensure the CompactDAQ chassis meets published specifications.
- Keep connections to the C Series module as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
- Maintain an ambient temperature of 23°C ± 5°C. The device temperature will be greater than the ambient temperature.
- Keep relative humidity below 90%.
- Allow a warm-up time of at least 10 minutes to ensure that the measurement circuitry is at a stable operating temperature.
Device Setup
- Install the C Series digital I/O module in the CompactDAQ chassis.
- Configure the hardware using Measurement & Automation Explorer (MAX).
- Launch the Calibration Executive procedure and complete the setup wizard. Ensure that you enter the CompactDAQ chassis serial number in the setup wizard and not the serial number of the C Series I/O module.