Measurement Rack Power States
- Updated2023-05-04
- 3 minute(s) read
Measurement Rack Power States
Power State | PEP | PDU | UPS | Temperature Controller |
---|---|---|---|---|
Off | Off | Off | Off | Off |
Standby | On (AC) | Off | Off | On |
EPO | DC only | Off | On | On |
Remote EPO | On (AC) | Off | On | On |
On | On (AC) | On | On | On |
Power State | Description | Action |
---|---|---|
Off | The system is entirely disabled with no power passing through the line filter or any internal test system components. | To enter Off state:
|
Standby | Power to the power distribution unit (PDU) and uninterruptible power supply (UPS) is shut off. Power is passing out of the PEP and into the thermostat emergency power off (EPO) panel. | To enter Standby state:
|
EPO | Immediately cuts AC power from the PEP and PDU. The UPS remains powered so that the
measurement instruments in the rack continue logging data. DC power
passes out of the PEP and into the thermostat EPO panel. Entering this state also triggers the remote EPO connector, which, when connected to an external safety loop, automatically shuts off external equipment. |
To enter EPO state, press the EPO button. To exit EPO state, turn clockwise to release the EPO button. |
Remote EPO | Immediately cuts power from the PDU. The UPS remains powered so that the measurement instruments in the rack continue logging data. Power passes out of the PEP and into the thermostat EPO panel. | The system enters Remote EPO state when other equipment
connected to an external safety loop through the remote EPO connector enter an EPO
state. The system exits Remote EPO state after the external instrument exits its EPO state. The rack switches into its previous state. |
On | A change to this state begins the main power on sequence of the test system. The PDU receives AC power through the UPS from the PEP and enables outlets to other system equipment. The EPO panel is powered. | To enter the On state, follow the process in Powering on the System. |