AIO-3682 Specifications

AIO-3682 Specifications

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Characteristics unless otherwise noted.

Conditions

Dual-output, first Nyquist zone specifications are valid under the following conditions unless otherwise noted.

  • Input DAC sampling clock: 2.5 GHz
  • Input power level: 1.5 dBm
  • Digital filter enable and selection
    • Register name: Data Mode Ctrl
      • Address: x18
      • Value: xE0
  • Full scale current register
    • Register name: FSC_1
      • Address: x20
      • Value: x00
    • Register name: FSC_2
      • Address: x21
      • Value: x01
  • Set operation mode register: baseband
    • Register name: Decode Ctrl
      • Address: x19
      • Value: x00

Dual-output, second Nyquist zone specifications are valid under the following conditions unless otherwise noted.

  • Input DAC sampling clock: 2.0 GHz
  • Input power level: 1.5 dBm
  • Digital filter enable and selection: Mix Mode
    • Register name: Data Mode Ctrl
      • Address: x18
      • Value: xE0
  • Digital filter enable and selection: baseband
    • Register name: Data Mode Ctrl
      • Address: x18
      • Value: xC0
  • Full scale current register
    • Register name: FSC_1
      • Address: x20
      • Value: x02
  • Set operation mode register: baseband
    • Register name: Decode Ctrl
      • Address: x19
      • Value: x00
  • Set operation mode register: Mix Mode
    • Register name: Decode Ctrl
      • Address: x19
      • Value: x01

General

Number of channels

2 DACs, single-ended, simultaneously sampled

Connector

SMA

Output impedance

50 Ω, per connector

Sample rate

1,400 MSPS to 2,850 MSPS

DAC part number

AD9129[1]1 For additional information on the DAC5682Z, refer to the DAC5682Z 16-Bit, 1.0 GSPS 2x-4x Interpolating Dual-Channel Digital-to-Analog Converter (DAC) data sheet at www.ti.com.; 14-bit resolution

Dual Output, First Nyquist Zone

AC-Coupled

Output range (normal operating conditions)

-0.5 dBm (0.6 Vpk-pk)

Bandwidth (-3 dB)

900 MHz

Frequency range

0.1 MHz to 900 MHz

SFDR (70 MHz Out, 2.5 GS/s, no PLL)

Non-harmonic

75 dBc

Second harmonic

56 dBc

Third harmonic

58 dBc

Channel-to-channel isolation

1 MHz

95 dB

10 MHz

100 dB

50 MHz

97 dB

100 MHz

100 dB

251 MHz

99 dB

500 MHz

100 dB

700 MHz

94 dB

Spectral Measurements (Dual Output, First Nyquist Zone)

Figure 1. Output Bandwidth (Baseband Mode)


Figure 2. 1.0 MHz Tone Spectral Measurement (100 MHz Bandwidth)


Figure 3. 10.0 MHz Tone Spectral Measurement (100 MHz Bandwidth)


Figure 4. 70.0 MHz Tone Spectral Measurement (600 MHz Bandwidth)


Figure 5. 150.0 MHz Tone Spectral Measurement (1,200 MHz Bandwidth)


Figure 6. 900.0 MHz Tone Spectral Measurement (5,000 MHz Bandwidth)


Dual Output, Second Nyquist Zone

AC-Coupled

Output range (normal operating conditions)

-0.5 dBm (0.6 Vpk-pk)

Table 1. Bandwidth (-3 dB)
Bandwidth Frequency Range
Baseband 850 MHz 150 MHz to 1,000 MHz
Mix Mode 1,750 MHz 1,000 MHz to 2,750 MHz
SFDR (70 MHz Out, 2 GS/s, no PLL)

Non-harmonic

76 dBc

Second harmonic

61 dBc

Third harmonic

72 dBc

Channel-to-channel isolation

150.4 MHz

92 dB

251.2 MHz

95 dB

500 MHz

98 dB

700 MHz

99 dB

1,250.4 MHz

93 dB

1,500 MHz

95 dB

2,500 MHz

78 dB

2,700 MHz

78 dB

Spectral Measurements (Dual Output, Second Nyquist Zone)

Figure 7. Output Bandwidth (Baseband Mode)


Figure 8. Output Bandwidth (Mix Mode)


Figure 9. 70.0 MHz Tone Spectral Measurement (2,000 MHz Bandwidth)


Figure 10. 70.0 MHz Tone Spectral Measurement (4,000 MHz Bandwidth)


Figure 11. 210.4 MHz Tone Spectral Measurement (0.1 MHz Bandwidth)


Figure 12. 210.4 MHz Tone Spectral Measurement (1.0 MHz Bandwidth)


Figure 13. 210.4 MHz Tone Spectral Measurement (20.0 MHz Bandwidth)


Figure 14. 210.4 MHz Tone Spectral Measurement (100.0 MHz Bandwidth)


Figure 15. 300.0 MHz Tone Spectral Measurement (2,000 MHz Bandwidth)


Figure 16. 300.0 MHz Tone Spectral Measurement (4,000 MHz Bandwidth)


Figure 17. 700.0 MHz Tone Spectral Measurement (2,000 MHz Bandwidth)


Figure 18. 700.0 MHz Tone Spectral Measurement (4,000 MHz Bandwidth)


Figure 19. 1,789.6 MHz Tone Spectral Measurement (0.1 MHz Bandwidth)


Figure 20. 1,789.6 MHz Tone Spectral Measurement (1.0 MHz Bandwidth)


Figure 21. 1,789.6 MHz Tone Spectral Measurement (20.0 MHz Bandwidth)


Figure 22. 1,789.6 MHz Tone Spectral Measurement (100.0 MHz Bandwidth)


Figure 23. 2,210.4 MHz Tone Spectral Measurement (0.1 MHz Bandwidth)


Figure 24. 2,210.4 MHz Tone Spectral Measurement (1.0 MHz Bandwidth)


Figure 25. 2,210.4 MHz Tone Spectral Measurement (20.0 MHz Bandwidth)


Figure 26. 2,210.4 MHz Tone Spectral Measurement (100.0 MHz Bandwidth)


CLK IN

General Characteristics

Number of channels[2]2 The second input is only necessary when using both DAC outputs.

2, single-ended

Connector

SMA

Input impedance

50 Ω

Input coupling

AC

External Sample Clock

Input voltage range

800 mVp-p to 1,575 mVpk-pk

Input frequency range

0.71 GHz to 2.85 GHz

Absolute maximum input

-0.3 V to 2.1 V

Maximum Power Requirements

Note Power requirements are dependent on the adapter modules installed and the contents of the FPGA application.

Power supply

3.3 V

Current

760 mA (2.5 GSPS), 850 mA (5.7 GSPS)

Physical

Dimensions (not including connectors)

7.65 cm × 6.91 cm × 0.68 cm (3.01 in. × 2.72 in. × 0.27 in.)

Weight

0.05 kg (0.11 lb)

Maximum Working Voltage

Note Maximum working voltage refers to the signal voltage plus the common-mode voltage.

Channel-to-earth

0 V to 3.3 V, Measurement Category I

Channel-to-channel

0 V to 3.3 V, Measurement Category I

Caution Do not use this device for connecting to signals in Measurement Categories II, III, or IV.

Environment

Maximum altitude 2,000 m (800 mbar) (at 25 °C ambient temperature)
Pollution Degree 2

Indoor use only.

Operating Environment

Operating temperature

-20 °C to 85 °C

Relative humidity range

10% to 90%, noncondensing (tested in accordance with IEC 60068-2-56)

Compliance and Certifications

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA C22.2 No. 61010-1
Note For safety certifications, refer to the product label or the Product Certifications and Declarations section.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
  • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-003: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the Product Certifications and Declarations section.
Notice To ensure the specified EMC performance, operate this product only with shielded cables and accessories.

CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:

  • 2014/35/EU; Low-Voltage Directive (safety)
  • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
  • 2011/65/EU; Restriction of Hazardous Substances (RoHS)
  • 2014/53/EU; Radio Equipment Directive (RED)
  • 2014/34/EU; Potentially Explosive Atmospheres (ATEX)

Product Certifications and Declarations

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

EU and UK Customers

  • Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
  • 电子信息产品污染控制管理办法(中国RoHS)

  • 中国RoHSNI符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于NI中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
  • 1 For additional information on the DAC5682Z, refer to the DAC5682Z 16-Bit, 1.0 GSPS 2x-4x Interpolating Dual-Channel Digital-to-Analog Converter (DAC) data sheet at www.ti.com.

    2 The second input is only necessary when using both DAC outputs.