Session to Refnum
Owning Palette: TestStand
Converts a viSession handle from an environment other than LabVIEW into a LabVIEW IVI or VISA refnum. For the Refnum parameter of the Refnum to Session VI and the Session to Refnum VI, pass only a VISA or IVI refnum created using the LabVIEW VISA Open function or the IVI New Session function. Passing an invalid refnum or a refnum constant can lead to undefined behavior and is not supported.
Details
|
Class Type specifies the IVI or VISA class of the underlying viSession handle you wire to the Session parameter. The function uses this parameter for type information only. |
|
Session specifies a viSession handle from an environment other than LabVIEW. |
|
error in (no error) describes error conditions that occur before this node runs. This input provides standard error in functionality. |
|
Refnum returns the LabVIEW IVI or VISA refnum for the underlying viSession handle Session . |
|
error out contains error information. This output provides standard error out functionality. |
Details
In a TestStand sequence, you can use a VI to create a LabVIEW IVI or VISA refnum, use the Refnum to Session function to convert the refnum to a viSession handle, and store the viSession handle in a TestStand variable. You can use the viSession handle in VIs that subsequent steps in a sequence call. You must use the Session to Refnum function to convert the viSession handle back to a LabVIEW IVI or VISA refnum. Close the session when you no longer need the refnum.
You cannot use a viSession or a LabVIEW IVI or VISA refnum in a process other than the process in which you created it.