Performing System Self-Calibration
- Aktualisiert2025-07-24
- 6 Minute(n) Lesezeit
Use the RTG Calibration Utility to calibrate the RTG.
To give users the ease of creating or modifying calibration files, RTG now comes packaged with a Calibration Utility. The Calibration Utility provides access to the required calibration parameters to be used within RTG. The Calibration Utility can be run through the main RTG application user interface or independently as a separate application.
Currently the Calibration Utility only supports local RTG systems.
The following figure depicts the RTG Calibration Utility state machine.
The states are as follows:
An RTG operates accurately when it is calibrated in the region of its operation. An RTG runs with a fixed Center Frequency and Reference Level; these two parameters define a unique calibration data set. A calibration file can contain as many calibration data sets as are needed to cover the expected range of operation.
The process to create a calibration data set is as follows:
-
Create a loopback connection between the output and input of the test fixture
as shown in the following figures.
Figure 40. Test Fixture Loopback Connection
Figure 41. Attenuation Sources
-
Launch the RTG Calibration Utility in one of the following ways:
- Run RTG Calibration Utility.exe from the <Program Files>\National Instruments\RTG\calibration-utility folder on the host computer.
- From Windows, select .
- From the RTG UI, select Calibration from the Utility menu.
The RTG Calibration Utility opens.
Figure 42. RTG Calibration Utility
Instrument Name Field Center Frequency Field Reference Level Field External LO Disabled/Enabled Button LO In Frequency Field LO In Power Field Calibration Utility Status Indicator Recommended LO In Frequency Indicator Recommended LO In Power Indicator Start Calibration Button Abort Calibration Button Minimum Latency Indicator Minimum Common Attenuation Indicator Maximum Common Attenuation Indicator Calibration Result Indicator Calibration Status Indicator - Select the VST to be calibrated from the Instrument Name field.
- Enter the Center Frequency to be calibrated.
-
Enter the value for the Reference Level to be
calibrated.
Refer to Reference Level and Pulse Detection Calibration for more information about configuring this field.
-
If you are connected to an external Local Oscillator (LO), configure the
following:
- Click the External LO Disabled button to enable.
-
Enter the LO In Frequency value (to the RTG center frequency or to a
different frequency within the upconverter/downconverter passband) in
the LO In Frequency field.
Note NI recommends setting the LO In Frequency to the RTG center frequency during calibration.The RTG software returns the expected LO In Frequency value based upon the VST instrument and center frequency in the Recommended LO In Frequency indicator.
-
Enter the LO In Power value.
The RTG software returns an expected LO In power based upon the mid-point value of the input range for a specific VST model in the Recommended LO In Power indicator.
-
Click Start Calibration to begin the calibration
process.
Note The process takes several minutes to complete.Note You can click Abort Calibration to return to the Inactive state before calibration completes. During some phases of system self-calibration, clicking Abort Calibration may not result in immediately moving RTG to the Idle state. The RTG Calibration Utility receives the Abort Calibration call, but waits until the current stage of the calibration to complete before moving to Idle.
While calibration is executing, the following occurs:
- Software updates the calibration status and calibration utility status on the RTG Calibration Utility UI.
When calibration completes, the following occur:- A calibration data set is written to the calibration file.
- The Calibration Utility UI displays the calibration results for the Minimum Latency, Minimum Common Attenuation, Maximum Common Attenuation, and Calibration Result.
- The Calibration Utility UI returns to the Idle state.
- Repeat steps 1 through 7 for all VSTs to be calibrated.
-
After calibration is complete, do one of the following:
- Calibrate with a new set of parameters.
- Exit the RTG Calibration Utility.
Saving and Using Stored Calibration Data
Running the RTG Calibration Utility saves data so that future RTG sessions with the same Center Frequency and Reference Level settings (as well as Data Path and Coprocessor Minimum Latency if using a coprocessor) can operate accurately.
RTG self-calibration features the following default behavior:
- Running the RTG Calibration Utility that returns a
PASSED
Calibration Result saves the calibration data
to disk.Note Run the Data Clearing Utility to remove any previously saved calibration data. Refer to Using the Data Clearing Utility for more information.
Verwandte Inhalte
- Using an External LO
Users are responsible for supplying an external LO based upon their application needs and requirements.
- PXIe-5830 Features
- PXIe-5831 User Manual
- PXIe-5832 Getting Started
- PXIe-5841 User Manual
- PXIe-5842 User Manual
- Reference Level and Pulse Detection Calibration
It is recommended that the radar signal levels be adjusted to a ±10 dBm range.
- Initializing Instruments and Opening a New Session in the RTG Software
Initializing moves RTG from the Idle state to the Inactive state.
- Using the Data Clearing Utility
Run this utility either when the application is shut down or no RTG applications are running.