PXIe-4082 Specifications

Version:
    Last Modified: February 21, 2017

    These specifications apply to the PXIe-4082.

    Definitions

    Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

    Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

    • Typical specifications describe the expected performance met by a majority of the models.
    • Nominal specifications describe parameters and attributes that may be useful in operation.

    Specifications are Warranted unless otherwise noted.

    Textcal is the device temperature at last external calibration.

    Tselfcal is the device temperature at last self-calibration.

    Conditions

    Specifications are valid under the following conditions unless otherwise noted. Refer to each section for additional conditions that apply.

    • Self-calibration performed within the last 24 hours
    • Calibration interval of 2 years
    • 60 minutes warm-up time

    DC Voltage Specifications

    Accuracy

    All DC voltage accuracy specifications apply to apertures of ≥100 ms, with Auto Zero and ADC calibration enabled. Assumes offset nulling. Otherwise, add 2 µV to the specifications.

    Table 1. DC Voltage ± (ppm of reading + ppm of range)
    Range Input Resistance[1] 24 Hr[2] Tselfcal ±1 °C 90 Day Tselfcal ±5 °C 2 Year Tselfcal ±5 °C Tempco/°C
    Without Self-Cal With Self-Cal
    100 mV 10 MΩ ± 2%, >10 GΩ 10 + 10 40 + 20 45 + 20 4 + 5 0.3 + 0.3
    1 V 6 + 2 20 + 6 25 + 6 2 + 1 0.3 + 0.3
    10 V 4 + 2 20 + 6 25 + 6 1 + 1 0.3 + 0.3
    100 V 10 MΩ ± 2% 6 + 2 30 + 6 35 + 6 4 + 1 0.3 + 0.3
    300 V 6 + 6 30 + 20 35 + 20 4 + 1 0.3 + 0.3

    Noise

    Figure 1. DC Voltage Noise, Typical
    spd-note-note
    Note  

    With input shorted, Normal DC Noise Rejection, and Auto Zero ON. For apertures less than 100 ms, add five times the typical rms noise to the accuracy specification.

    General

    ADC Linearity

    0.5 ppm of reading + 1 ppm of range

    Effective Common-Mode Rejection Ratio (CMRR) (1 kΩ resistance in LO lead)

    >140 dB (DC), 100 ms aperture; >170 dB (>46 Hz) with high-order DC noise rejection, 100 ms aperture, typical

    Overrange

    105% of range except 300 V

    DC voltage input bias current

    <30 pA at 23 °C, typical

    Resistance Specifications

    Accuracy

    All resistance accuracy specifications apply to apertures of ≥100 ms, with Offset Compensated Ohms (for ranges ≤10 kΩ) or Auto Zero (for ranges ≥ 100 kΩ) and ADC calibration enabled.

    Table 2. Resistance (4-Wire and 2-Wire [3]) ± (ppm of reading + ppm of range)
    Range Test Current[4] Max Test Voltage 24 Hr[5] Tselfcal ± 1 °C 90 Day Tselfcal ± 5 °C 2 Year Tselfcal ± 5 °C Tempco/°C
    Without Self-Cal With Self-Cal
    100 Ω 1 mA 100 mV 15 + 10 50 + 15 80 + 15 5 + 1 0.8 + 1
    1 kΩ 1 mA 1 V 12 + 2 50 + 3 80 + 3 5 + 0.1 0.8 + 0.1
    10 kΩ 100 µA 1 V 12 + 2 50 + 3 80 + 3 5 + 0.1 0.8 + 0.1
    100 kΩ[6] 10 µA 1 V 15 + 2 90 + 6 95 + 6 5 + 0.5 2 + 0.5
    1 MΩ 10 µA 10 V 20 + 2 90 + 10 95 + 10 5 + 1 2 + 1
    10 MΩ 1 µA 10 V 100 + 2 800 + 10 800 + 10 20 + 3 20 + 3
    100 MΩ[7] 1 µA || 10 MΩ 10 V 500 + 10 3000 + 10 3000 + 10 300 + 10 300 + 10

    Noise

    Figure 2. PXIe-4082 Resistance Noise, Typical
    spd-note-note
    Note  

    With input shorted, Normal DC Noise Rejection, and Auto Zero ON. For apertures less than 100 ms, add five times the typical rms noise to the accuracy specification.

    General

    Maximum 4-wire lead resistance

    Use the lesser of 10% of range or 1 kΩ

    DC Current Specifications

    Accuracy

    All DC current accuracy specifications apply for apertures ≥100 ms, with Auto Zero and ADC calibration enabled.

    Table 3. DC Current ± (ppm of reading + ppm of range)
    Range Burden Voltage, Typical 24 hour[8] Tselfcal ± 1 °C 2 Year Tselfcal ± 5 °C Tempco/°C
    20 mA <20 mV 20 + 15 450 + 200 8 + 10
    200 mA <200 mV 20 + 15 550 + 20 8 + 1
    1 A <800 mV 20 + 15 700 + 50 8 + 2

    Noise

    Figure 3. PXIe-4082 DC Current Noise, Typical
    spd-note-note
    Note  

    With input open, Normal DC Noise Rejection, and Auto Zero ON. For apertures less than 100 ms, add five times the typical rms noise to accuracy specification.

    General

    Overrange

    105% of range except 1 A range.

    AC Voltage Specifications

    Accuracy

    spd-note-note
    Note  

    Measurement aperture greater than 4/fL where fL is the lowest frequency component of the signal being measured. Signal amplitudes greater than 1% of range.

    Table 4. AC Voltage Accuracy ± (% of reading + % of range), 2 Years, T extcal ± 10 °C, T selfcal ± 5 °C
    Range (rms) Peak Voltage 1 Hz to 40 Hz[9] >40 Hz to 20 kHz >20 kHz to 50 kHz >50 kHz to 100 kHz >100 kHz to 300 kHz
    50 mV[10] ±105 mV 0.1 + 0.04 0.05 + 0.04 0.09 + 0.04 0.5 + 0.08 3 + 0.1
    500 mV ±1.05 V 0.1 + 0.01 0.05 + 0.02 0.09 + 0.02 0.5 + 0.02 3 + 0.05
    5 V ±10.5 V
    50 V ±105 V
    300 V ±450 V
    Tempco/°C 0.001 + 0.001 0.001 + 0.001 0.001 + 0.001 0.001 + 0.001 0.01 + 0.01

    General

    Input impedance

    1 MΩ ± 2% in parallel with 150 pF, typical

    Input coupling

    AC or DC coupled

    Overrange

    105% of range except 300 V

    Maximum Volt-Hertz product

    Verified to 2.2 x 107 V-Hz

    Maximum DC voltage component

    250 V

    Common mode rejection ratio (CMRR), 1 kΩ resistance in LO lead

    >70 dB (DC to 60 Hz), typical

    AC Current Specifications

    Accuracy

    spd-note-note
    Note  

    Measurement aperture greater than 4/fL, where fL is the lowest frequency component of the signal being measured. Signal amplitudes greater than 1% of range.

    Table 5. AC Current Specifications ± (% of reading + % of range), 2 Years, Full operating temperature range
    Range (rms) Peak Current Burden Voltage (rms), Typical 1 Hz to 20 kHz[11] Tempco/°C
    10 mA ±20 mA <10 mV 0.04 + 0.02 0.001 + 0.0001
    100 mA ±200 mA <100 mV 0.04 + 0.02 0.001 + 0.0001
    1 A ±2 A <800 mV 0.1 + 0.02 0.001 + 0.0001

    General

    Overrange

    105% of range except 300 V

    Capacitance Specifications

    Accuracy Specifications

    Table 6. Capacitance ± (% of reading + % of range), 2 Years, T extcal ± 10 °C
    Range (rms) Accuracy[12] Tempco/°C Effective Test Current,[13] Nominal Effective Frequency,[13] Nominal Default Model Maximum Reading Rate[14]
    300 pF 0.5 + 0.6 0.01 + 0.025 160 nA 3 kHz Parallel 15 S/s
    1 nF 0.4 + 0.2 0.01 + 0.003 330 nA 3 kHz Parallel 15 S/s
    10 nF 0.3 + 0.1 0.01 + 0.001 330 nA 3 kHz Parallel 15 S/s
    100 nF 0.3 + 0.1 0.01 + 0.001 3.3 µA 3 kHz Parallel 15 S/s
    1 µF 0.3 + 0.1 0.01 + 0.001 100 µA 1 kHz Series 15 S/s
    10 µF 0.3 + 0.1 0.01 + 0.001 1 mA 1 kHz Series 15 S/s
    100 µF 0.3 + 0.1 0.01 + 0.001 1 mA 91 Hz Series 3 S/s
    1000 µF 0.4 + 0.1 0.01 + 0.001 1 mA 91 Hz Series 3 S/s
    10000 µF 0.3 + 0.1 0.01 + 0.001 1 mA 91 Hz Series 3 S/s

    General Specifications

    DC bias[15]

    0.46 V from HI to LO, nominal, user-selectable (OFF by default)

    Inductance Specifications

    Accuracy Specifications

    Table 7. Inductance ± (% of reading + % of range), 2 Years, T extcal ± 10 °C
    Range (rms) Accuracy[16] Tempco/°C Effective Test Current,[17] Nominal Effective Frequency,[17] Nominal Default Model Maximum Reading Rate[18]
    10 µH 0.5 + 1 0.01 + 0.01 330 µA 30 kHz Series 20 S/s
    100 µH 0.5 + 0.1 0.01 + 0.01 330 µA 30 kHz Series 20 S/s
    1 mH 0.5 + 0.1 0.01 + 0.0001 330 µA 3 kHz Series 15 S/s
    10 mH[19] 0.5 + 0.1 0.005 + 0.001 3.3 µA 3 kHz Series 15 S/s
    100 mH[19] 0.5 + 0.1 0.005 + 0.001 33 µA 273 kHz Series 3 S/s
    1 H[19] 0.5 + 0.1 0.007 + 0.001 3.3 µA 273 kHz Series 3 S/s
    5 H[19] 0.5 + 0.1 0.007 + 0.001 330 nA 273 kHz Series 3 S/s

    Diode Test Specifications

    Range

    10 V

    Test current[20]

    1 µA, 10 µA, 100 µA, 1 mA[21]

    Accuracy

    Add 20 ppm of reading to 10 VDC voltage specifications.

    Frequency and Period Specifications

    spd-note-note
    Note  

    Aperture time set to 150 ms.

    Frequency range

    15 Hz to 500 kHz

    Period measurement range

    2 µs to 66.67 ms

    AC Input Voltage Range Corresponding Isolated Digitizer Range Minimum Peak-to-Peak Signal Amplitude[22] Maximum Peak-to-Peak Signal Amplitude Accuracy
    50 mV 100 mV 5 mV 200 mV Refer to the PXIe_CLK100 accuracy of the chassis.
    500 mV 1 V 50 mV 2 V
    5 V 10 V 500 mV 20 V
    50 V 100 V 5 V 200 V
    300 V 300 V 50 V 450 V

    Temperature Specifications

    All temperature accuracy specifications apply to apertures ≥100 ms, Auto Zero, and ADC calibration enabled. Use lowest possible resistance or voltage range for each temperature. Add probe accuracy and cold junction accuracy where applicable.

    Sensor Type Temperature Range Accuracy
    RTD[23] -200 to 600 °C 0.1 °C
    Thermistor[24] -80 to 150 °C 0.08 °C
    J Thermocouple -210 to 1200 °C 0.2 °C
    K Thermocouple -200 to 1200 °C 0.3 °C
    N Thermocouple -200 to 1300 °C 0.4 °C
    T Thermocouple -200 to 400 °C 0.3 °C
    E Thermocouple -200 to 1000 °C 0.2 °C
    R Thermocouple -50 to 1760 °C 0.8 °C
    S Thermocouple -50 to 1760 °C 0.8 °C
    B Thermocouple 400 to 1820 °C 0.8 °C

    Isolated Digitizer Specifications

    Available functions

    Voltage and current

    Voltage ranges

    ±100 mV to ±300 V (DC or AC coupled)

    Current ranges

    ±20 mA to ±1 A

    Sample rate range

    10 S/s to 1.8 MS/s

    Available sample rates

    r = (1.8 MS/s) / y, where y = 1, 2, 3,...1.8 x 105

    Timebase accuracy

    Equal to the PXIe_CLK100 accuracy of the chassis

    Digitizer record length

    2 samples minimum, unlimited maximum

    Table 8. Voltage Mode
    Range Input Resistance[25] DC Accuracy, (ppm/reading + ppm/range) 2 year, Tselfcal ± 5 °C Analog Bandwidth, [26] Typical
    ±0.1 dB -3 dB
    100 mV 1 MΩ ± 2%, >10 GΩ 125 + 175 40 kHz 240 kHz
    1 V 125 + 75 40 kHz 240 kHz
    10 V 125 + 75 40 kHz 240 kHz
    100 V 1 MΩ ± 2% 125 + 75 30 kHz 240 kHz
    300 V 125 + 75 30 kHz 240 kHz
    Figure 4. PXIe-4082 Voltage Waveform Noise, Typical
    spd-note-note
    Note  

    With input shorted.

    Table 9. Current
    Range Burden Voltage, Typical DC Accuracy, (ppm/reading + ppm/range) 2 year, Tselfcal ± 5 °C Analog Bandwidth, Typical
    ±0.1 dB -3 dB
    20 mA <20 mV 100 +100 60 kHz 300 kHz
    200 mA <200 mV 100 +100 60 kHz 300 kHz
    1 A <800 mV 100 +100 60 kHz 300 kHz
    Figure 5. PXIe-4082 Current Waveform Noise, Typical
    spd-note-note
    Note  

    With input open.

    General Specifications

    External calibration interval

    2 years

    Warm-up

    60 minutes to rated accuracy

    Measurement Category

    II

    spd-note-caution
    Caution  

    Do not use this device for connection to signals or for measurements within Measurement Categories III or IV.

    Input protection (between terminals or terminal to ground)

    300 VDC or ACrms

    Current mode fuse

    T 1 A 400 V, time-lag user-replaceable Minimum interrupt rating: 500 A Littelfuse 0477001.MXP

    Maximum common-mode voltage

    300 VDC or ACrms

    Maximum voltage to earth ground

    HI

    300 VDC or ACrms

    LO

    300 VDC or ACrms

    HI SENSE

    300 VDC or ACrms

    LO SENSE

    300 VDC or ACrms

    spd-note-fuse-symbol
    Fuse  

    When this fuse symbol is marked on a device, take proper precautions.

    spd-note-
    Hazardous Voltage  

    This icon denotes a warning advising you to take precautions to avoid electrical shock.

    Timing

    Mode Trigger Latency Maximum Reading Rate[27]
    Voltage, current, and resistance ≤556 ns 20 kS/s
    Voltage and current digitizer
    Capacitance and inductance 30 ms, nominal Refer to capacitance and inductance specifications for maximum reading rates.

    Power

    Power consumption

    <9 W from PXI Express backplane

    +12 V load

    0.55 A max

    + 3.3 V load

    0.55 A max

    Physical Characteristics

    Dimensions

    3U, one-slot, PXI/cPCI module; 2.0 cm x 13.0 cm x 21.6 cm (0.8 in. x 5.1 in. x 8.5 in.), nominal

    Weight

    340 g (12 oz), nominal

    spd-note-note
    Note  

    If you need to clean the device, wipe it with a dry towel.

    Environment

    Maximum altitude

    2,000 m (800 mbar) (at 25 °C ambient temperature)

    Pollution Degree

    2

    Indoor use only.

    Operating Environment

    Ambient temperature range

    0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.)

    Relative humidity range

    10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Storage Environment

    Ambient temperature range

    -40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.)

    Relative humidity range

    5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

    Shock and Vibration

    Operating shock

    30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.)

    Random vibration

    Operating

    5 Hz to 500 Hz, 0.3 grms (Tested in accordance with IEC 60068-2-64.)

    Nonoperating

    5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

    Compliance and Certifications

    spd-note-caution
    Caution  

    Electromagnetic interference can adversely affect the measurement accuracy of this product. The input terminals of this device are not protected for electromagnetic interference. As a result, this device may experience reduced measurement accuracy or other temporary performance degradation when connected cables are routed in an environment with radiated or conducted radio frequency electromagnetic interference. To limit radiated emissions and to ensure that this device functions within specifications in its operational electromagnetic environment, take precautions when designing, selecting, and installing measurement probes and cables.

    Safety

    This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

    • IEC 61010-1, EN 61010-1
    • UL 61010-1, CSA C22.2 No. 61010-1
    spd-note-note
    Note  

    For UL and other safety certifications, refer to the product label or the Online Product Certification section.

    Electromagnetic Compatibility

    This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
    • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
    • EN 55011 (CISPR 11): Group 1, Class A emissions
    • EN 55022 (CISPR 22): Class A emissions
    • EN 55024 (CISPR 24): Immunity
    • AS/NZS CISPR 11: Group 1, Class A emissions
    • AS/NZS CISPR 22: Class A emissions
    • FCC 47 CFR Part 15B: Class A emissions
    • ICES-001: Class A emissions
    spd-note-note
    Note  

    In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.

    spd-note-note
    Note  

    Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.

    spd-note-note
    Note  

    For EMC declarations, certifications, and additional information, refer to the Online Product Certification section.

    CE Compliance

    This product meets the essential requirements of applicable European Directives, as follows:

    • 2014/35/EU; Low-Voltage Directive (safety)
    • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)

    Online Product Certification

    Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.

    Environmental Management

    NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

    For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

    Waste Electrical and Electronic Equipment (WEEE)

    spd-note-weee
    EU Customers  

    At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.

    电子信息产品污染控制管理办法(中国RoHS)

    spd-note-china-rohs
    中国客户  

    National Instruments符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于National Instruments中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)

    • 1 In parallel with 150 pF, typical
    • 2 Relative to external calibration source.
    • 3 Perform offset nulling or add 200 mΩ to reading.
    • 4 -10% to 0% tolerance, typical.
    • 5 Relative to external calibration source.
    • 6 Perform offset nulling or add 2 ppm of range to the specifications.
    • 7 2-wire resistance measurement only. Use tempco outside Textcal +/- 10 °C. Typical accuracy is 5% between 105 MΩ and 1.05 GΩ.
    • 8 Relative to external calibration source.
    • 9 Applies to DC coupled only.
    • 10 Applies to signals >2 mV
    • 11 Specification is typical for the 5 kHz to 20 kHz frequency range.
    • 12 After lead compensation with <3 meters of coaxial or shielded twisted-pair cabling. Specifications apply to >5% of range and <110% of range, except 300 pF range which measures down to 0.05 pF.
    • 13 Correlated to single-tone test method.
    • 14 Number of LC measurements to average = 1
    • 15 Applies to capacitance modes only.
    • 16 After lead compensation with <3 meters of coaxial or shielded twisted-pair cabling. Specifications apply to >5% of range and <110% of range, except 300 pF range which measures down to 0.05 pF.
    • 17 Correlated to single-tone test method.
    • 18 Number of LC Measurements to Average = 1.
    • 19 Specifications apply to >1% of range.
    • 20 -10% to 0% tolerance, typical.
    • 21 Up to 4.5 V measurement for 1 mA test current.
    • 22 Square wave input. Minimum required peak-to-peak signal level is valid only for frequencies up to the -3 dB bandwidth. For higher frequencies, the signal amplitude must be increased. Refer to the Digitizer Voltage Mode for bandwidths.
    • 23 Based on Pt3851 RTD in a 4-wire configuration.
    • 24 Based on 44004, 44006, and 44007 interchangeable thermistors.
    • 25 Input impedance in parallel with 150 pF, typical. When AC coupled, only 1 MΩ available.
    • 26 Typical AC coupled frequency is 6 Hz (±0.1 dB) and 0.8 Hz (-3 dB).
    • 27 Maximum Reading Rate assumes minimum aperture time, Auto Zero is OFF, Offset Compensated Ohms is OFF, ADC Calibration is OFF, Number of Averages is 1, and Settle Time is 0 seconds. Varying these settings will vary the reading rate.