For the NI 5102, this node externally calibrates a specified channel and vertical range.
Call this node before calling Adjust Range Calibration to perform a preliminary offset calibration. The first time this node is called, a default vertical range parameter is used in the driver to calibrate the vertical offset. When this node is called again after Adjust Range Calibration, it performs a precise offset calibration using the calibrated vertical range.
The instrument handle that you obtain from Initialize External Calibration. The handle identifies a particular instrument session.
Error conditions that occur before this node runs.
The node responds to this input according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Default: No error
The vertical range, in volts, to calibrate.
Reference to the NI-SCOPE instrument session to pass to the next node in the program.
Error information.
The node produces this output according to standard error behavior.
Standard Error Behavior
Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.
Where This Node Can Run:
Desktop OS: Windows
FPGA: Not supported