Self Calibrate (G Dataflow)

Version:
    Last Modified: March 3, 2017

    Self-calibrates most NI digitizers, including all SMC-based devices and most Traditional NI-DAQ (Legacy) devices.

    connector_pane_image
    datatype_icon

    session in

    Identifies a particular instrument session.

    datatype_icon

    error in

    Error conditions that occur before this node runs.

    The node responds to this input according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Default: No error

    datatype_icon

    channels

    The channel to calibrate.

    Channel String Syntax

    This input parameter has the following options:

    • A single channel, such as 0.
    • A list of channels, such as 0,1 or 3,2,1,0
    • A range of channels, such as 0-7 or 0:7.
    • All channels, which is designated by an empty string.
    spd-note-note
    Note  

    The order of channels in a list is important. The attributes are set on the channels in the order they occur in the string.

    datatype_icon

    option

    Self-calibrates all channels or restores the external calibration.

    Name Value Description
    Self Calibrate All Channels 0 Performs a self-calibration on all available channels.
    Restore External Calibration 1 Restores the external calibration.
    datatype_icon

    session out

    Passes the instrument handle to the next node in the program. This parameter has the same value as session in.

    datatype_icon

    error out

    Error information.

    The node produces this output according to standard error behavior.

    Standard Error Behavior

    Many nodes provide an error in input and an error out output so that the node can respond to and communicate errors that occur while code is running. The value of error in specifies whether an error occurred before the node runs. Most nodes respond to values of error in in a standard, predictable way.

    error in does not contain an error error in contains an error
    If no error occurred before the node runs, the node begins execution normally.

    If no error occurs while the node runs, it returns no error. If an error does occur while the node runs, it returns that error information as error out.

    If an error occurred before the node runs, the node does not execute. Instead, it returns the error in value as error out.

    Self-Calibrating SMC-Based Devices

    For SMC-based devices, if the self-calibration is performed successfully in a regular session, the calibration constants are immediately stored in the self-calibration area of the EEPROM. If the self-calibration is performed in an external calibration session, the calibration constants take effect immediately for the duration of the session. However, they are not stored in the EEPROM until you call Cal End with action set to Store Calibration and no errors occur.
    spd-note-note
    Note  

    Not all digitizers support self-calibration.

    Where This Node Can Run:

    Desktop OS: Windows

    FPGA: Not supported