DAQmx Setup SCXI Calibration (1502) (G Dataflow)

Last Modified: September 18, 2017

Sets an SCXI-1502 module to the specified gain value for calibration.

spd-note-note
Note  

The terminal where you can measure module output depends on the configuration of the module in MAX. National Instruments recommends cabling the module to the digitizer, so that the output appears on the MCH0± pins of the rear signal connector.

connector_pane_image
datatype_icon

calHandle in

calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.

datatype_icon

physical channel

physical channel specifies the physical channel to calibrate.

datatype_icon

gain

gain specifies the gain setting to calibrate.

datatype_icon

error in

error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

datatype_icon

calHandle out

calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.

datatype_icon

error out

error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.