DAQmx Setup FieldDAQ Calibration (11637) (G Dataflow)

Last Modified: August 15, 2018

Writes the specified binary value to the DAC on the specified physical channel.

This module requires a sweep of DAC values obtained from the DAQmx Get 9269 Calibration Adjustment Points node.

connector_pane_image
datatype_icon

calHandle in

calHandle in is a reference to the calibration session that you created using the DAQmx Initialize External Calibration node.

datatype_icon

physical channels

physical channels specifies the name of the physical channel to use to create the virtual channel. The DAQmx physical channel constant lists all physical channels on devices and modules installed in the system.

You also can wire a string that contains a physical channel to this input. You cannot specify more than one physical channel.

datatype_icon

bridgeConfig

bridgeConfig specifies the Wheatstone bridge configuration connected to the channel.

Name Description
Full Bridge

Four active sensing elements in the bridge.

Half Bridge

Two active sensing elements in the bridge.

Quarter Bridge 350 Ohm Completion Resistor

One active sensing element with a 350 ohm completion resistor.

Quarter Bridge 120 Ohm Completion Resistor One active sensing element with a 120 ohm completion resistor.
datatype_icon

excitationValue

excitationValue specifies in volts the amount of excitation supplied to the sensor. Refer to the sensor documentation to determine appropriate excitation values.

datatype_icon

error in

error in describes error conditions that occur before this node runs. This input provides standard 'error in' functionality.

datatype_icon

calHandle out

calHandle out is a reference to the calibration session. Wire this output to other external calibration nodes.

datatype_icon

error out

error out contains error information. If error in indicates that an error occurred before this node ran, error out contains the same error information. Otherwise, error out describes the error status that this node produces.