Filtered Results

Specifications listed in examples are for demonstration purposes only and do not necessarily reflect specifications for this device. Example 1: Determining Extended Range Pulse On Time and Duty Cycle Parameters ...

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, ...

These specifications apply to the PXIe-4139. In this document, the PXIe-4139 (40W) and PXIe-4139 (20W) are referred to inclusively as the PXIe-4139. The information in this document applies to all versions ...

The following table and figures illustrate the voltage and the current source and sink ranges of the PXIe-4139. Current Source and Sink Ranges DC voltage ranges DC current source and sink ranges 600 mV ...

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity EN ...

This product is intended for use in indoor applications only. Cover all empty slots using filler panels.

This product meets the requirements of the following environmental standards for electrical equipment. IEC 60068-2-1 Cold IEC 60068-2-2 Dry heat IEC 60068-2-78 Damp heat (steady state) IEC 60068-2-64 Random ...

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment ...

Temperature Operating 0 °C to 55 °C Storage -40 °C to 71 °C Humidity Operating 10% to 90%, noncondensing Storage 5% to 95%, noncondensing Pollution Degree 2 Maximum altitude 2,000 m (800 mbar) (at 25 °C ...

Specifications listed in examples are for demonstration purposes only and do not necessarily reflect specifications for this device. Example 1: Calculating 5 °C Accuracy Calculate the accuracy of 900 nA ...

VoltageDevice configured for local sense100 μV per mA of output load change (measured between output channel terminals), typical Device configured for remote sense Load regulation effect included in voltage ...

Available sample ratesWhen sourcing while measuring, both the Source Delay and Aperture Time affect the sampling rate. When taking a measure record, only the Aperture Time affects the sampling rate. (1.8 ...

Wideband source noise<20 mV peak-to-peak in 60 V range, device configured for normal transient response, 10 Hz to 20 MHz, typical The following figures illustrate measurement noise as a function of measurement ...

Output Resistance Programming Accuracy Current Level/Limit Range Programmable Resistance Range, Voltage Mode Programmable Resistance Range, Current Mode Accuracy ± (% of resistance setting), T[cal] ± 5 ...

Dimensions 3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.) Weight PXIe-4139 (40W) 427 g (15.1 oz) PXIe-4139 (20W) 419 g (14.8 oz) Front panel ...

PXI Express power requirement PXIe-4139 (40W) 3.0 A from the 3.3 V rail and 6.0 A from the 12 V rail PXIe-4139 (20W) 2.5 A from the 3.3 V rail and 2.2 A from the 12 V rail

Output channel protection Overcurrent or overvoltageAutomatic shutdown, output disconnect relay opensOvertemperatureAutomatic shutdown, output disconnect relay opens

Dynamic load, minimum pulse cycle timeFor example, given a continuous pulsin load, if the largest dynamic step in power that the load sources/sinks is from 5 W to 15 W, then the maximum SMU power step ...

Showing 1-20 of 108 results