The following item is the bug fix in PXIe-6571 18.0.0f2.
ID | Legacy ID | Fixed Issue | Issue Details |
---|---|---|---|
714493 |
| PXIe-6571 devices may become inoperable after self-calibration. | After performing self-calibration on the PXIe-6571 the device may become inoperable. Once the device becomes inoperable the only fix is to RMA the device. |