NI Calibration Executive 3.6.3跑PXIe-6674T校准程序时常产生错误,这无法忽略并且阻止了将新的调整数据写入设备。为了解决该问题,需要使用附件中的文件来替换如下位置中的VI.
32位操作系统: C:\Program Files\National Instruments\Calibration Executive\procedures\NI 6674T\Support Files\Adjust\...
64位操作系统: C:\Program Files (x86)\National Instruments\Calibration Executive\procedures\NI 6674T\Support FIles\Adjust\...
|