The following items are notable issues fixed between the release of NI-RFSA 2025 Q4 and NI-RFSA 2026 Q1, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of NI-RFSA.
|
Bug Number |
Legacy ID |
Description |
Details |
|---|---|---|---|
| 3593690 |
Incorrect measurements with NI-RFSA/NI-RFSG 2025 Q4 and PXIe-5633 running 1.1.1 firmwareUsing NI-RFSA 2025 Q4 or NI-RFSG 2025 Q4 with PXIe-5633 modules running firmware 1.1.1 or later causes incorrect two-port S-parameter and wave measurements. Workaround: Upgrade to NI-RFSA 2026 Q1 or later and NI-RFSG to 2026 Q1 or later.
|
Reported Version: NI-RFSA: 2025 Q4 | NI-RFSG: 2025 Q4 Resolved Version: NI-RFSA: 2026 Q1 Added: Dec 9, 2025 |
Installing some patches may require certain additional steps or considerations. Please refer to the following table for more information about patches for this release.
These patches currently do not have any special instructions.