The following items are notable issues fixed between the release of NI-DCPower 2026 Q3 and NI-DCPower 2026 Q4, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of NI-DCPower.
|
Bug Number |
Legacy ID |
Description |
Details |
|---|---|---|---|
| 3946295, 3979570, 4000729, 4000739 |
Self calibration on the PXIe-4135/6/7 might cause bad calibration, especially in a system under sufficient loadOn PXIe-4135, PXIe-4136, and PXIe-4137 devices, performing niDCPower Self Calibrate while the system is under heavy load or experiencing high latency can result in degraded calibration accuracy.
Workaround: Perform self-calibration on modules serially, while minimizing system load and latency, to maximize calibration accuracy.
|
Reported Version: NI-DCPower: 20.6 Resolved Version: NI-DCPower: 2026 Q4 Added: N/A |
Installing some patches may require certain additional steps or considerations. Please refer to the following table for more information about patches for this release.
These patches currently do not have any special instructions.