The following items are notable issues fixed between the release of NI-SCOPE 2024 Q1 and NI-SCOPE 2024 Q3, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of NI-SCOPE.
|
Bug Number |
Legacy ID |
Description |
Details |
|---|---|---|---|
| 2774430, 1478240, 373083 |
Degradation of DC Accuracy when using non-standard sample clock timebase rates with PXIe-5172 devices in driver software releases from 2022 Q4 through 2023 Q4An improvement to the Self Calibration of the PXIe-5172 was added in NI-SCOPE 21.0. This only impacts users who use the NISCOPE_ATTR_SAMP_CLK_TIMEBASE_RATE attribute and set it to either 150 MS/s or 200 MS/s. This improvement was mistakenly removed from several versions, including NI-SCOPE 2023 Q1, NI-SCOPE 2023 Q3, and NI-SCOPE 2023 Q4. It was re-added starting in NI-SCOPE 2024 Q1.
Workaround: Install a version not impacted by the regression.
|
Reported Version: LabVIEW Instrument Design Libraries for Reconfigurable Oscilloscopes 2022 Q4 | LabVIEW Instrument Design Libraries for Reconfigurable Oscilloscopes 2023 Q2 | NI-SCOPE 2023 Q1 | NI-SCOPE 2023 Q3 | NI-SCOPE 2023 Q4 Resolved Version: NI-SCOPE 2024 Q1 Added: Jul 9, 2024 |