This document contains the NI-FGEN known issues that were discovered before and since the release of NI-FGEN 2024 Q4. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.
Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community; see Additional Resources.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
227853 | 624206 |
Reset button not available in MAX or Hardware Configuration Utility for PXIe-5413/5423/5433 waveform generatorsWorkaround: Use the niFgen Reset Device VI or niFgen_ResetDevice function. |
Reported Version: NI-FGEN 17.1 Resolved Version: N/A Added: May 8, 2017 |
227788 | 631405 |
Common MAX features are not supported on PXIe-5413/5423/5433 waveform generators: Test Panels, Device Pinouts, Device RoutesWorkaround: Refer to the NI Signal Generators Help for pinout and routes documentation for PXIe-5413/5423/5433 waveform generators. |
Reported Version: NI-FGEN 17.1 Resolved Version: N/A Added: May 8, 2017 |
227806 | 642679 |
PFI lines on PXIe-5450 and PXIe-5451 waveform generators drive a short pulse when NI-FGEN initialize functions are called with the reset device parameter set to FalseWorkaround: Set the reset device parameter to True when initializing an NI-FGEN session to a PXIe-5450 and PXIe-5451 waveform generator. |
Reported Version: NI-FGEN 17.1 Resolved Version: N/A Added: May 8, 2017 |
227817 | 637708 |
niFgen ni5433 External Calibration API is not translated to Japanese in LabVIEWWorkaround: Refer to the NI Signal Generators Help for Japanese documentation of the niFgen ni5433 External Calibration API. |
Reported Version: NI-FGEN 17.1 Resolved Version: N/A Added: May 8, 2017 |
227814 | 632366 |
Initializing and closing an NI-FGEN session in a loop leaks memory for NI-DAQmx-Based Waveform GeneratorsWorkaround: There is currently no known workaround for this issue. |
Reported Version: NI-FGEN 15.0 Resolved Version: N/A Added: May 8, 2017 |
227755 | 644381 |
Japanese characters not allowed in waveform names or script names on PXIe-5413/5423/5433 waveform generatorsNI-DAQmx-based waveform generators allow waveform names with Japanese characters, but on PXIe-5413/5423/5433 waveform generators the NI-FGEN API will report error -1074101561 (0xBFFA82C7): "The script contains an invalid character or symbol. Replace the invalid character with a valid symbol or alphanumeric character." Workaround: Change waveform names to use only characters from the English alphabet. |
Reported Version: NI-FGEN 17.1 Resolved Version: N/A Added: May 8, 2017 |
227744 | 687100 |
Synchronization between the PXIe-5413/5423/5433 and the PXIe-5442 does not work in Standard Function output mode and with the PXIe-5442 as the master deviceWorkaround: To synchronize the PXIe-5442 and a PXIe-5413/5423/5433 in Standard Function output mode, use the PXIe-5413/5423/5433 as the master device or read the TClk period and then set the minimum TClk period to a higher value by calling niTClk Synchronize. |
Reported Version: NI-FGEN 18.1 Resolved Version: N/A Added: May 17, 2018 |
227830 | 641019 |
Device data underflow failure occurs when generating waveforms with a size of 196 or 200 in Arbitrary Sequence mode on the PXI-5422 waveform generatorThe PXI-5422 product documentation states that the minimum supported waveform size is 192 samples and that larger values that are multiples of 4 are also supported. However, a device data underflow error occurs when using waveform sizes of 196 and 200. Workaround: Double the number of samples on the host and double the sample rate. |
Reported Version: NI-FGEN 15.0 Resolved Version: N/A Added: May 8, 2017 |
537276 | 643582 |
Write waveform function times out when streaming on PXIe-5413/5423/5433 waveform generators if the sample rate is below 256 kS/s or if the sample rate is below 4.5 MS/s and the allocated waveform size is below 1.2 k samplesWorkaround: Duplicate samples on the host and increase the sample rate of the waveform generator accordingly. |
Reported Version: NI-FGEN 17.1 Resolved Version: N/A Added: May 8, 2017 |
537291 | 642393 |
Options String "DriverSetup=MemorySize:" doesn't simulate memory size correctly on PXIe-5413/5423/5433 waveform generatorsThe value passed for the MemorySize tag in the Driver Setup string is ignored in simulated instrument sessions. The simulation logic assumes 128 MB of usable memory per channel for PXIe-5413 and PXIe-5423 waveform generators, and 512 MB of usable memory per channel for PXIe-5433 waveform generators.
Workaround: There is currently no known workaround for this issue. |
Reported Version: NI-FGEN 17.1 Resolved Version: N/A Added: May 8, 2017 |
2818196 |
PXIe-5413/23/33 Arb Sequence stepped trigger has poor alignment, compared to initial start trigger, varying as the arb sample rate is changedThe delay for a stepped trigger on the PXIe-5413, PXIe-5423, and PXIe-5433 to occur increases and has more jitter the lower the sample rate. Workaround: There is currently no known workaround for this issue. |
Reported Version: NI-FGEN 17.1 | NI-FGEN 2024 Q3 Resolved Version: N/A Added: Sep 26, 2024 |
|
1210589, 2256290 |
NI-FGEN devices may fail under Linux kernel DMA remappingAttempts to use NI-FGEN devices may fail on Linux kernels that have enabled DMA remapping. The Linux kernel logs will contain errors similar to the following: Linux distributions that enable DMA remapping by default include Ubuntu 22.04 with kernel 6.8 or higher and Ubuntu 24.04.
Workaround: Modify the Linux kernel DMA remapping configuration as outlined in https://www.ni.com/r/iommu-linux.
|
Reported Version: NI-FGEN: 21.3 Resolved Version: N/A Added: Sep 27, 2024 |
Issues found in this section will not be listed in future known issues documents for this product.
There are currently no issues to list.