The following items are notable issues fixed between the release of NI-DCPower 2024 Q3 Patch 3 and NI-DCPower 2024 Q4, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of NI-DCPower.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
2758054 |
On the PXIe-4135/6/7 when Output Cutoff is triggered, it cannot be disabled without calling resetting the deviceWhen Output Cutoff is enabled on the NI PXIe-4135, PXIe-4136, PXIe-4137, if a condition is reached which activates the cutoff on a channel, calling Reset with Channels will not clear the output cutoff condition. In this situation, the channel will continue to output a near-zero voltage and current.
Workaround: Call Reset Device to remove the Output Cutoff condition.
|
Reported Version: NI-DCPower 2024 Q2 Resolved Version: NI-DCPower 2024 Q4 Added: N/A |
|
2561819 |
An internal hardware error (-235140) intermittently occurs on PXIe-4135/6/7On the NI PXIe-4135, PXIe-4136 and PXIe-4137, in certain situations, these instruments may return an internal hardware error (-235140) with the Self Test Code 268435457 in the error description. Workaround: Perform a device reset on the instrument.
|
Reported Version: NI-DCPower 20.6 Resolved Version: NI-DCPower 2024 Q4 Added: N/A |
|
2558394, 2657571, 2683830, 2707843, 2780294, 2803670, 2805273, 2855482 |
NI-DCPower Fetch and Measure may cause a system crash under very high CPU load when using the PXIe-4147, PXIe-4162, or PXIe-4163When retrieving measurements from the PXIe-4147, PXIe-4162, or PXIe-4163, a race condition may cause a system crash (Blue Screen of Death). The likelihood of the crash increases with an increased number of simultaneous calls to niDCPower Fetch Multiple or niDCPower Measure Multiple (for example, when using a Parallel For Loop in LabVIEW). Workaround: Serialize calls to niDCPower Fetch Multiple and niDCPower Measure Multiple. Use niDCPower Wait for Event to wait for the Measure Complete Event prior to calling Fetch Multiple.
|
Reported Version: NI-DCPower 21.8 Resolved Version: NI-DCPower 2024 Q3 Patch 3 Added: N/A |
Installing some patches may require certain additional steps or considerations. Please refer to the following table for more information about patches for this release.
These patches currently do not have any special instructions.