Calibration Executive 6.9 Known Issues

Overview

This document contains the Calibration Executive known issues that were discovered before and since the release of Calibration Executive 6.9. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.

 

Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community; see Additional Resources.

Bug Number

Legacy ID

Description

Details

208070 677002

Procedure Throws Error or Terminates When a PXIe-4080/4081/4082 DMM Is Used

PXIe-408x digital multimeters are not compatible with several Calibration Executive procedures for which a PXI-407x digital multimeter is recommended as an instrument standard.

For more information and a list of procedures, refer to "DMM Instrument Standard Considerations" in Calibration Executive Help.

Workaround:

Do not substitute a PXIe-4080/4081/4082 in a procedure in which a PXI-4070/4071/4072 DMM is recommended.

Reported Version:

Calibration Executive 4.2

Resolved Version:

N/A

Added:

Aug 14, 2020

1090478

NIPM Does Not Enforce Dependencies in Calibration Executive Feed

NI Package Manager allows the Calibration Executive feed to be modified to remove core components that are required for certain procedures. Dependencies should not be modified or deselected during installation.

Workaround:

There is currently no known workaround for this issue.

Reported Version:

Calibration Executive 6.0

Resolved Version:

N/A

Added:

Aug 14, 2020

1094770

Error Occurs Generating a Calibration Report on a Non-English OS

Windows OS for languages other than English (Spanish, for example) may use a date format with extended characters. In this case, PDF, Word, and Excel reports cannot be generated using default report options; Calibration Executive displays an error:
The report file was not at the expected path or has been corrupted.

Workaround:

Configure report options to remove Calibration Date from the report file name. For details about customizing report names, refer to "Configuring Report Options" in Calibration Executive Help.

Reported Version:

Calibration Executive 6.0

Resolved Version:

N/A

Added:

Aug 14, 2020

1091448

Installing Version 6.0 on Japanese OS with Suggested Components Causes Error

When you install Calibration Executive 6.0 on a Japanese operating system and accept the pre-selected additional items, an error occurs at the NI-HSDIO-Documentation package.

Workaround:

When you install Calibration Executive on a Japanese OS, deselect all NI-HSDIO components from the additional items list in NI Package Manager. In this case, however, the Japanese OS system cannot be used to calibrate high-speed digital I/O hardware.

Reported Version:

Calibration Executive 6.0

Resolved Version:

N/A

Added:

Oct 8, 2020

1090572

Error Occurs If User Imports Custom Report Configuration from a Different Folder

From the Custom Report Configurator, if you try to import a new report configuration from a location other than the default folder, the following error occurs:
The import was unsuccessful!
Please check if the selected folder is a correct report configuration.

Workaround:

Copy individual stylesheet folders to the default report configuration folder location. In the Custom Report Configurator, click Import, and select the main CustomReports folder to import all stylesheets, or select each stylesheet folder individually.

Reported Version:

Calibration Executive 6.0

Resolved Version:

N/A

Added:

Dec 14, 2020

1084777

Error Occurs If Path to Custom Report Stylesheet or CustomReports Folder Changes

If the file path for a custom report stylesheet or the CustomReports folder changes (for example, if a user's Documents folder is cleaned up), Calibration Executive displays an error when the report generator is launched:
Calibration Executive:  Report Gen: Report stylesheets, templates, or schema are not available. Reinstall Calibration Executive to restore these necessary files.
C:\Users\<username>\Documents\CalibrationExecutive\CustomReports\cex_CustomReports_config.ini
The report generator then closes abruptly.

Workaround:

Do one of the following:
  • Copy the content of the CustomReports folder (including the style sheet) to the expected location.
  • Manually edit the cex_Config.ini file at C:\Program Files\National Instruments\Calibration Executive\Config\, and change CustomReportsPath to point to the existing location (if known).
  • Delete the CustomReportsPath line in the cex_Config.ini file to force Calibration Executive to reset to the default location.

Reported Version:

Calibration Executive 6.0

Resolved Version:

N/A

Added:

Aug 14, 2020

1065569

Custom Report Features Cannot Be Shared Across User Accounts on a Workstation

Access to custom reports on a single workstation is limited to the account used to install and use Calibration Executive. By default, the path for custom reports data is:
C:\Users\<username>\Documents\CalibrationExecutive\CustomReports

Workaround:

Copy the content of the CustomReports folder (including the style sheet) to a shared folder, and then edit the cex_Config.ini file at C:\Program Files\National Instruments\Calibration Executive\Config\ to change CustomReportsPath to the new location.

Reported Version:

Calibration Executive 6.0

Resolved Version:

N/A

Added:

Dec 14, 2020

Final Time Issue Listed

Issues found in this section will not be listed in future known issues documents for this product.

There are currently no issues to list.

Additional Resources

Explore Support Content and Product Documentation 

 

 

Ask the NI Community

 

 

Request Support from an Engineer

 

A valid service agreement may be required, and support options vary by country

 

Glossary of Terms

 

  • Bug ID - When an issue is reported to NI, you may be given this ID or find it on ni.com.  You may also find IDs posted by NI on the discussion forums or in KnowledgeBase articles.
  • Legacy ID – An older issue ID that refers to the same issue.  You may instead find this issue ID in older known issues documents.
  • Description - A few sentences which describe the problem. The brief description given does not necessarily describe the problem in full detail.
  • Workaround - Possible ways to work around the problem.
  • Reported Version - The earliest version in which the issue was reported.
  • Resolved Version - Version in which the issue was resolved or was no longer applicable. "N/A" indicates that the issue has not been resolved.
  • Date Added - The date the issue was added to the document (not the reported date).