ni.com is currently experiencing issues.
Support teams are actively working on the resolution.
ni.com is currently experiencing issues.
Support teams are actively working on the resolution.
The following items are notable issues fixed between the release of NI-DCPower 2022 Q3 and NI-DCPower 2022 Q4, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of NI-DCPower.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
2065379 |
PXIe-4162 or PXI-4163 may return error -1074097799 when performing a device resetOccasionally, resetting a PXIe-4162 or PXI-4163 SMU will return error -1074097799 (The device session is invalid). After this occurs, all subsequent calls to the instrument will return error -1074116989 (Internal Hardware Error) until the device is reset again. Workaround: Reset the device using NI MAX, the NI-DCPower API, or the System Configuration API: Alternatively, upgrade to NI-DCPower 2022 Q4. |
Reported Version: NI-DCPower 20.7 | NI-DCPower 21.0 | NI-DCPower 21.3 | NI-DCPower 21.3.1 | NI-DCPower 21.5 | NI-DCPower 21.8 | STS Software Bundle: 20.0 | STS Software Bundle: 20.1 | STS Software Bundle: 21.0 | STS Software Bundle: 21.0.1 | STS Software Bundle: 22.0 Resolved Version: NI-DCPower 2022 Q4 Added: N/A |
Installing some patches may require certain additional steps or considerations. Please refer to the following table for more information about patches for this release.
These patches currently do not have any special instructions.