This document contains the Battery Test System Software known issues that were discovered before and since the release of Battery Test System Software 2022 Q4. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.
Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community; see Additional Resources.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
1596248 |
Test Monitor does not attach large test reports to SystemLinkTestStand report files over 80 MB are not uploaded to SystemLink and attached to the corresponding test result.
Workaround: Ensure your report files are smaller than 80 MB or transfer them manually. Complete the following steps to ensure the report is unique to each DUT if you are using the batch process model:
|
Reported Version: Battery Test System Software 2.0 Resolved Version: N/A Added: Sep 1, 2021 |
|
1873086 |
Battery test package files cannot be removed once submitted through the BatteryTestPackageLoaderOnce battery tests package files are submitted through the BatteryTestPackageLoader, they cannot be removed from the database of the server.
Workaround: Submitting a package file with the same name updates the existing configuration. In this case, the old version is archived and used only by existing test plans. Submitting a configuration with a different name results in a new test plan.
|
Reported Version: Battery Test System Software 2022 Q2 Resolved Version: N/A Added: Jun 27, 2022 |
|
1985043 |
The NI Battery Test System Web Server does not start properly in slow computersThe NI Battery Test System Web Server windows service does not start properly and displays a timeout error in slow computers.
Workaround: Increase the Windows service timeout value following the instructions here: A slow service does not start due to time-out error in Windows
|
Reported Version: Battery Test System Software 2022 Q2 Resolved Version: N/A Added: Jun 27, 2022 |
|
2168173 |
Embedded Data Logger settings are ignored when using Size in the base propertyThe BTS Server ignores the segmenting settings for EDL.xml files that use "Size" as their Base, such as in: <FileSegmenting Interval="60" Base="Size" Enabled="true"/>
Workaround: Use "Time" instead of "Size" for the Base in the FileSegmenting element of the EDL.xml, such as in: <FileSegmenting Interval="60" Base="Time" Enabled="true"/>
|
Reported Version: Battery Test System Software 2022 Q2 Resolved Version: N/A Added: Oct 24, 2022 |
|
2168457 |
Test station files must not have more than one NI-VCOM (PROVEtechRBS) instrumentA test station file (.teststation) with more than one NI-VCOM (PROVEtechRBS) instrument will be accepted by the package loader application and will make the test station not work properly.
Workaround: Do not add more than one NI-VCOM (PROVEtechRBS) instrument in a test station file (.teststation).
|
Reported Version: Battery Test System Software 2022 Q2 Resolved Version: N/A Added: Oct 24, 2022 |
|
2170408 |
Package loader cannot intake packages due to an error writing to the EntityFileAsset.index fileIntaking test packages through the package loader application may, in some cases, result in the following error: "The process cannot access the file '\\?\C:\bts_production_storage\EntityFileAsset.index' because it is being used by another process."
Workaround: Run the package loader command again.
|
Reported Version: Battery Test System Software 2022 Q2 Resolved Version: N/A Added: Oct 24, 2022 |
|
2174258 |
The Embedded Data Logger and SystemLink custom devices display error 1162 when opening their right-click menusThe Embedded Data Logger and SystemLink custom devices display Error 1162: "LabVIEW: (Hex 0x48A) Cannot find one or more tags" when opening their right-click menus in the VeriStand System Explorer. Workaround: Use the icons in the top bar of the Veristand System Explorer for most of the right-click menu functionality. If needed, temporarily relocate or remove the following files when working with the configuration of these custom devices in a VeriStand system definition file and then restart VeriStand:
After restarting, return the files to their original locations after accessing the required right-click menu options and closing VeriStand.
|
Reported Version: Battery Test System Software 2022 Q2 Resolved Version: N/A Added: Oct 24, 2022 |
|
2174304 |
Some DUT Ports are not published to SystemLink as tagsSome NI-VCOM (PROVEtechRBS) channels are not published to SystemLink as tags.
Workaround: List the NI-VCOM (PROVEtechRBS) instrument before the SystemLink instrument in the test station configuration file (.teststation).
|
Reported Version: Battery Test System Software 2022 Q2 Resolved Version: N/A Added: Oct 24, 2022 |
Issues found in this section will not be listed in future known issues documents for this product.
There are currently no issues to list.