The following items are notable issues fixed between the release of Calibration Executive 6.0 and Calibration Executive 6.1, including additional patches and service packs. If you have an issue ID, you can search this list to validate that the issue has been fixed. This is not an exhaustive list of issues fixed in the current version of Calibration Executive.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
1082441 |
Error Occurs During PXIe-5645R Adjust Only OperationWhen the PXIe-5654R calibration procedure is run for Adjust Only, the following error occurs:
Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.0 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |
|
1046555 |
USB-5132/5133 Test Points Fail "AC Amplitude Accuracy" StepUSB-5132 and USB-5133 low voltage test points fail in the AC Amplitude Accuracy step due to ground loop noise. Workaround: Make sure the ground wire on the back of the DUT is connected to the signal source ground. |
Reported Version: Calibration Executive 6.0 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |
|
1082203 |
PDF Report Is Missing Columns If Results Do Not Fit on PageWhen a PDF calibration report that displays several test configuration parameters also displays Uncertainty, the As-Left column might be truncated, leaving off the PassFail column. See the Downloads section for an example. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 4.6 Resolved Version: Calibration Executive 6.1 Added: Dec 8, 2020 |
|
988160 |
PXI-4110 Fails "Verify Current Line Regulation" and Other TestsThe PXI-4110 module may falsely fail Verify Current Line Regulation and other tests because the output takes longer than the time provided to settle below specified limits. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 5.1 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |
|
1134882 |
Procedure Displays Generic Error -300401, Calibration Results Are EmptyWhen an error in a procedure occurs before the Caldata30 VI captures any data, then the process model issues an error message that is too generic to be useful for debugging the error:
Workaround: On the Tests tab, locate the test step that triggered the error, and hover the cursor over the "Error" status. The tooltip displays detailed information for the actual error. |
Reported Version: Calibration Executive 6.0 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |
|
1103134 |
PXIe-5654 Procedure Unable to Verify RF Out Amplitude AccuracyIn the PXIe-5654 calibration procedure, "RF OUT Amplitude Accuracy" verifies test points to which not all units can be adjusted. This issue can cause "false" test failures. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 5.0 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |
|
1179187 |
PXIe-5665 (3.6 GHz) -nopreamp Throws Error at "SpecAn Characterization"The PXIe-5665 (3.6 GHz) -nopreamp procedure throws an error at the beginning of the SpecAn Characterization step. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.0.1 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |
|
208090 |
NI 54xx Procedures Do Not Retry Failed Flatness Test Reference PointsFor NI 54xx calibration procedures, when a Flatness test reference point fails, the procedure does not retry the test point as expected. If the reference value is bad, all subsequent flatness test points are invalid. This issue affects the following variants:
Workaround: Ensure connections are correct before proceeding with the Flatness test. |
Reported Version: Calibration Executive 4.2 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |
|
451713 |
PXIe-5654 ZeroAndSelfCalibrate Function Throws Error for Unsupported InstrumentsDuring the "Zero and Calibrate the Power Meter" step, the PXIe-5654 and PXIe-5654 with PXIe-5696 calibration procedures (all variants) throw an error for unsupported instruments. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 5.0 Resolved Version: Calibration Executive 6.1 Added: Dec 3, 2020 |