Understanding NI PXI DMM Architecture for High-Accuracy and Flexible Measurements​

Contents

Overview

The NI PXI Digital Multimeter (DMM) combines measurement accuracy, speed, software integration, and modular PXI system flexibility across a range of performance tiers. The NI PXI DMM includes options with the NI FlexDMM architecture, a measurement architecture designed to support precision digital multimeter functionality alongside advanced analysis capabilities traditionally requiring additional instrumentation. The highest-performance implementation of this architecture is found in the NI PXIe-4081 7½-digit PXI DMM, and the same architectural foundations are leveraged across all NI PXIe-408x DMMs. This paper explores the measurement conversion technologies, architectural decisions, and subsystem designs that enable the performance, flexibility, and long-term measurement stability of the PXIe-408x PXI DMMs.

​What Is the FlexDMM Architecture? 

The term FlexDMM refers to NI modular PXI digital multimeters that combine precision measurement capability with software-defined measurement processing and advanced analysis workflows. Unlike DMM architectures optimized for certain measurement functions, the FlexDMM architecture was designed to support a broad range of measurement and analysis workflows within a PXI system. The FlexDMM architecture serves as the foundation for capabilities such as precision voltage, current, and resistance measurements; advanced self-calibration; software-based signal analysis; and waveform acquisition on supported models.

What Is FlexADC Technology?

​At the heart of the FlexDMM architecture is the FlexADC, the NI measurement conversion architecture. FlexADC combines multiple analog-to-digital conversion techniques to balance competing design objectives including measurement precision, linearity, throughput, stability, and waveform acquisition capability. Rather than optimizing for a single performance attribute, the FlexADC was designed to support a broad range of measurement applications within a common architecture. To better understand this design approach, it is useful to examine the measurement conversion architectures commonly used in precision digital multimeters.

Measurement Conversion Architectures in Precision Digital Multimeters

​Different digital multimeter architectures make different trade-offs between resolution, speed, complexity, noise performance, and flexibility. Over time, several conversion approaches have emerged as common solutions within precision measurement instruments. Understanding these approaches helps explain the design decisions behind the FlexADC architecture used in the PXIe-408x DMMs . 

Dual-Slope ADC Technology

Dual-slope converters represent one of the oldest and most widely used approaches to precision analog-to-digital conversion. The technique integrates an unknown input signal over a fixed period and then compares that accumulated charge against a known reference. This architecture offers excellent noise rejection and high measurement accuracy, making it well-suited for precision DMM measurements. However, because the conversion process relies on sequential integration and discharge cycles, measurement throughput is fundamentally limited. As measurement resolution increases, conversion times typically increase as well. Figure 1 illustrates a simplified dual-slope converter architecture.

Dual-Slope Converter Block Diagram

Dual-Slope Converter Block Diagram

Dual-Slope Converter Block Diagram

Figure 1. Dual-Slope Converter Block Diagram
 

This technique is used in many high-resolution DMMs, even today. It has the advantage of simplicity and precision. With long integration times, you can increase resolution to theoretical limits. However, the following design limitations ultimately affect product performance:

  • Dielectric absorption of the integrator capacitor must be compensated, even with high-quality integrator capacitors, which can require complicated calibration procedures.
  • The signal must be gated on and off, as must the reference. This process can introduce charge injection into the input signal. Charge injection can cause input-dependent errors (nonlinearity), which are difficult to compensate for at very high resolutions (6½ digits or more).
  • The ramp-down time seriously degrades the speed of measurement. The faster the ramp-down, the greater the errors introduced by comparator delays, charge injection, and so on.


Some topologies use a transconductance stage prior to the integrator to convert the voltage to a current and then use “current steering” networks to minimize charge injection. Unfortunately, this added stage introduces complexity and possible errors.

Despite these design limitations, dual-slope converters have been used in DMMs from the most common bench or field service tools to high-precision, metrology-grade, high-resolution DMMs. As with most integrating A/D techniques, they have the advantage of providing fairly good noise rejection. Setting the integration period to a multiple of 1/PLC (power line frequency) causes the A/D to reject line frequency noise—a desirable result.

Charge-Balance with Ramp-Down ADC Technology

Charge-balance converters build upon the principles of dual-slope conversion by continuously applying reference corrections throughout the integration process. This approach reduces some of the limitations associated with traditional dual-slope converters while improving measurement throughput. Charge-balance architectures have become widely adopted in precision measurement instruments because they provide a practical balance between accuracy and speed. However, they still fundamentally operate as integrating converters and remain optimized for discrete measurement results rather than continuous waveform acquisition. Figure 2 illustrates a simplified charge-balance converter architecture.

Simplified charge-balance converter architecture block diagram
 
 
Simplified charge-balance converter architecture block diagram
 
Figure 2. Charge-Balance Converter Block Diagram


During the integration phase, represented in Figure 2 by taperture, S1 is turned on and Vx is applied through R1, which starts the integrator ramping. Opposing current is applied at regular intervals through switches, S2 and S3. This “balances” the charge on C1. Measurement counts are generated each time S5 is connected to VR. In fact, for higher-resolution measurements (longer integration times), most of the counts are generated during this taperture phase. At the end of the charge-balance phase, a precision reference current is applied to the integrator, as is done in the case of the dual-slope converter. The integrator is thus ramped down until it crosses zero. The measurement is calculated from the counts accumulated during the integration and added to the weighted counts accumulated during the ramp-down. Manufacturers use two or more ramp-down references, resulting in fast ramp-downs to optimize speed and then slower “final slopes” for precision.

Although you can greatly improve your integrator capacitor dielectric absorption problems with the charge-balance with ramp-down A/D, it has performance benefits similar to the dual-slope converter. (In fact, some dual-slope converters use multiple ramp-down slopes.) Speed is greatly improved because the number of counts generated during the charge-balance phase reduces the significance of any ramp-down error, so ramp-down can be much faster. However, there is still significant dead time if you make multiple measurements or if you digitize a signal because of disarming and rearming the integrator.

Sigma-Delta Converter Technology

Sigma-delta converters use oversampling, digital filtering, and noise shaping techniques to achieve high effective resolution and excellent linearity. These characteristics have made sigma-delta converters the foundation of many modern precision measurement instruments and data acquisition systems. Unlike integrating converter architectures, sigma-delta converters are inherently continuous-sampling systems and can provide substantial flexibility in how measurement data is processed. However, sigma-delta designs typically require trade-offs among bandwidth, latency, resolution, and filtering behavior depending on application requirements. Figure 3 illustrates a simplified sigma-delta converter architecture..

Simplified sigma-delta converter architecture block diagram

Figure 3. Sigma-Delta Converter Block Diagram


The basic building blocks of a sigma-delta converter are the integrator or integrators, one-bit ADC and DAC (digital-to-analog converter), and digital filter. You conduct noise shaping by combining the integrator stages and digital filter design. Different philosophies exist regarding the optimum number of integrator stages, number of digital filter stages, number of bits in the A/D and D/A converters, and so on. However, the basic operational building blocks remain fundamentally the same. A modulator consisting of a one-bit charge-balancing feedback loop is similar to that described above. The one-bit ADC, because of its inherent precision and monotonicity, leads the way to very good linearity.

There are many advantages to using commercially available sigma-delta converters:

  • They are fairly linear and offer good differential nonlinearity (DNL).
  • You can control signal noise very effectively.
  • They are inherently self-sampling and tracking (no sample-and-hold circuitry required).

However, there are some limitations to using off-the-shelf sigma-delta ADCs in high-resolution DMMs:

  • Speed limitations, especially in scanning applications, due to pipeline delays through the digital filter
  • Precision limitations set by the manufacturer
  • Modulation “tones” can alias into the passband, creating problems at high resolutions
  • Limited control over speed-noise trade-offs, acquisition time, and so on

The FlexADC Architecture Behind NI PXI DMMs

The FlexADC architecture is the measurement engine used for NI PXIe-408x DMMs. Rather than relying exclusively on a single conversion approach, FlexADC combines high-speed acquisition technology with precision conversion techniques to provide a powerful architecture capable of supporting both traditional DMM measurements and advanced analysis workflows. The highest-performance implementation of this architecture can be found in the PXIe-4081, which combines 7½-digit measurement capability, expanded measurement ranges, and advanced waveform acquisition functionality. Other PXI DMMs leverage the same architectural principles while optimizing for different application requirements. Figure 4 provides a simplified conceptual view of the FlexADC architecture.

FlexADC Converter architecture block diagram with sigma-delta converter, feedback DAC, and digital signal processor (DSP)
Figure 4. FlexADC Converter


The block diagram in Figure 4 shows a simplified model of how the FlexADC operates. At low speeds, the circuit exploits the advantages of the sigma-delta converter. The feedback DAC is designed for extremely low noise and exceptional linearity. The lowpass filter provides the noise shaping necessary for effective performance across all resolutions. No ramp-down is needed because the ultra-high-precision 1.8 MS/s modulator provides extremely high-resolution conversion without it. At high speeds, the 1.8 MS/s modulator combined with the fast-sampling ADC, deliver continuous-sample digitizing. The digital signal processor (DSP) provides real-time sequencing, calibration, linearization, AC true-RMS computing, and decimation, as well as the weighted noise filtering used for the DC functions.

The FlexADC architecture enables several capabilities for NI PXI DMMs:

  • Configurable trade-offs between measurement resolution and throughput
  • Support for both precision measurements and waveform-based analysis
  • Continuous acquisition techniques that avoid interruption of the measurement signal path
  • Software-defined signal processing and analysis
  • Advanced digital filtering and noise-rejection techniques
  • Common architectural foundations across multiple PXI DMM performance tiers

​The unique architecture offers a continuously variable reading rate from 7 S/s at 7½ digits to 10 kS/s at 4½ digits, as shown in Figure 5.

 

Digits versus reading rate performance graph for NI PXIe-4081
Figure 5. PXIe-4081 DC Reading Rates


Table 1 compares the four previously described conversion architectures.

 


Table 1. Measurement Conversion Architecture Comparison

Precision Measurement Foundation

The performance of a precision digital multimeter depends not only on its conversion architecture, but also on the quality and stability of its measurement front end. The NI PXI DMM combines precision and highly stable voltage and resistance references, low-noise signal conditioning, and advanced self-calibration techniques to maintain measurement accuracy and long-term stability.

Solid-State Input Signal Conditioning

A major source of measurement error in most traditional DMMs is electromechanical relay switching. Contact-induced thermal voltage offsets can cause instability and drift. The FlexDMM architecture eliminates all but one relay in the DCV, ACV, and resistance path. A special relay-contact configuration cancels the thermal errors in this single relay. This relay is switched only during self-calibration. All measurement-related switching for function and range changing is done with low-thermal, highly reliable solid-state switching. Thus, electromechanical relay wear-out failures are all but eliminated. Figure 6 shows an overnight drift performance of the most sensitive range, the 100 mV range. Each division is 500 nV. For comparison, the same measurement made under identical conditions with a traditional 6½-digit DMM and a full-rack 8½-digit DMM is shown in Figure 6.

FlexDMM versus traditional DMM noise and draft graph. Curve shows FlexDMM (lower) range stability with shorted input, compared to a traditional DMM.

Figure 6. Curve Showing FlexDMM (lower) 100 mV Range Stability with Shorted Input, Compared to a Traditional DMM (upper) – 500 nV/Division

 

Linearity

Linearity is a measure of the “quality” of a DMM transfer function. It’s important in conversion-component characterization applications to offer differential nonlinearity (DNL) and integral nonlinearity (INL) performance substantially better than that available in off-the-shelf ADCs. The FlexADC is designed for excellent linearity, both DNL and INL. Linearity is also important because it determines the repeatability of the self-calibration function. Figure 7 shows a typical FlexDMM linearity plot measured on the 10 V range from -10 to +10 V.

Linearity plot measured on the 10 V range

Figure 7. 10 VDC Range Linearity

Self-Calibration

Precision measurements are influenced by temperature, component aging, and environmental conditions. To maintain measurement accuracy across changing operating conditions, NI PXI DMMs incorporate integrated self-calibration capability that characterizes and compensates for measurement-path errors using onboard precision references.

Self-calibration helps maintain measurement performance after temperature changes without requiring an external calibration procedure. This capability is especially important in PXI systems where instruments operate inside automated test racks, manufacturing environments, or other systems where ambient conditions may vary over time. Self-calibration can be helpful for following reasons:

  • Self-calibration corrects signal-path gain and offset errors back to the onboard precision voltage reference.
  • Resistance self-calibration corrects source, gain, and offset errors back to the onboard precision resistance reference.
  • Self-calibration supports maintaining specified performance across the operating temperature range for supported functions.

Having self-calibration available in NI PXI DMMs results in delivering a highly accurate, ultra stable digital multimeter in wide operating temperature range, well outside of the traditional 18 to 28 °C. Table 2 compares the accuracy of a traditional 7½-digit example DMM against the PXIe-4081.

 

Table 2. Example Summary: Uncertainty Analysis, Measuring 5 V on 10 V Range

Notice that using the FlexDMM architecture with self-calibration provides accuracy at 50 °C, which is more than seven times better than traditional methods. Table 3 compares self-calibration to the traditional “Factory” calibration.

 

Table 3. Calibration Comparison

Voltage Measurement Architecture

The voltage measurement architecture used throughout the NI PXI DMM was developed to support precision DC measurements, accurate AC measurements, and waveform acquisition capabilities on supported models within a compact PXI module. Achieving these goals requires careful consideration of signal conditioning, bandwidth, stability, isolation, and measurement linearity.

High-Voltage Architecture

On the PXIe-4081, high-voltage measurement capability requires careful consideration of component breakdown, voltage spacing, attenuator design, and power dissipation within the compact PXI form factor. These design considerations enable measurements up to the PXIe-4081 voltage limits while maintaining the spacing and safety requirements appropriate for the instrument category.

Single 10 MΩ Input Attenuator

Traditionally, DMMs use both 1 MΩ and 10 MΩ attenuators in their front-end signal conditioning circuits. A 1 MΩ attenuator, when used to provide the signal conditioning for a 700 VAC signal, dissipates more than 0.5 W. This requirement is difficult for a miniature precision component. Controlling temperature-coefficient-induced errors is a challenge even for a physically large component. Therefore, eliminating the need for the 1 MΩ attenuator is required. Another important reason why 1 MΩ attenuators are used in traditional DMMs is the wide AC bandwidth. Traditional 10 MΩ attenuator designs cannot attain the same level of AC bandwidth achieved by 1 MΩ attenuator designs.

The PXIe-4081 FlexDMM architecture incorporates an innovative scaled bootstrap design to null the attenuator capacitance that traditionally impedes wide bandwidth performance. This bootstrap, shown in Figure 8, is designed and carefully laid out to minimize stray capacitance loading the 100 kΩ attenuator leg of the input attenuator network RN. With the addition of the scaled bootstrap formed by R1–R4, C1, and U1, flat step response is assured. Most importantly, the characteristic response achieved is very close to that of a single pole RC, which is important for digitizer and DC step response.

FlexDMM architecture scaled bootstrap block diagram
Figure 8. PXIe-4081 Scaled Bootstrap

Additionally, the PXIe-4081 uses the digital AC DSP flatness correction to compensate the residual attenuator flatness without the use of compensation capacitors. These two compensation techniques deliver an order of magnitude improvement over what would otherwise be possible given the requirement that the single attenuator be capable of passing ACRMS, precision DC, and digitizer signals.

Component Breakdown and Voltage Spacing

One of the most daunting enemies to high-voltage measurement is range selection switch (relay) breakdown. Traditionally, DMMs use high-voltage relays. High-voltage relay switching and high reliability are not easy to achieve together in any package, let alone a miniaturized one. To meet both these requirements, the PXIe-4081 implements a novel new solid-state device for range selection capable of withstanding well over 1000 V in the off state. This device has none of the traditional reliability problems of electromechanical relays because there are no contacts to be damaged by high-voltage switching, and no contact life limitations. The secondary benefit of solid-state input signal conditioning is excellent low-level DC thermal performance.

By moving to solid-state high-voltage switching, eliminating the need for a 1 MΩ divider, and using DSP to eliminate calibration components, the PXIe-4081 meets the CAT I requirements that make measurements up to 1,000 V possible in PXI.

DC Noise Rejection

The FlexDMM architecture includes configurable digital noise-rejection techniques that allow engineers to optimize measurements for different electrical environments. By adjusting how measurement samples are weighted and processed, the instrument can improve rejection of interfering signals while maintaining measurement throughput.

Normal

When you select normal DC noise rejection, all samples are weighted equally. This process emulates the behavior of most traditional DMMs, providing good rejection of frequencies at multiples of f0 where f0 = 1/taperture , the aperture time selected for the measurement. Figure 9 illustrates normal weighting and the resulting noise rejection as a function of frequency. Notice that good rejection is obtained only very near multiples of f0.


Normal DC noise rejection graphs illustrating normal weighting and the resulting noise rejection as a function of frequency
Figure 9. Normal DC Noise Rejection

Second-Order

Second-order DC noise rejection applies a triangular weighting to the measurement samples, as shown in Figure 10. Notice that very good rejection is obtained near even multiples of f0, and that rejection increases more rapidly with frequency than with normal sample weighting. Also notice that the response notches are wider than they are with normal weighting, resulting in less sensitivity to slight variations in noise frequency. You can use second-order DC noise rejection if you need better power line noise rejection than you can get with normal DC noise rejection but you can’t afford to sample slowly enough to take advantage of high-order noise rejection. For example, you can set the aperture to 33.333 ms for a 60 Hz power line frequency.

Second-order DC noise rejection showing a triangular weighting to the measurement samples
Figure 10. Second-Order DC Noise Rejection

High-Order

Figure 11 illustrates high-order sample weighting and its resulting noise rejection as a function of frequency. Notice that noise rejection is good starting around 4f0 and is excellent above 4.5f0. Using high-order DC noise rejection, you achieve almost no sensitivity to noise at any frequency above 4.6f0. A FlexDMM architecture using high-order DC noise rejection with a 100 ms aperture (10 readings/s) can deliver full 6½-digit accuracy with more than 1 V of interfering power-line noise on the 10 V range at any frequency higher than 46 Hz. This is the equivalent of >110 dB normal mode rejection, insensitive to variations in power-line frequency.

High-order DC noise rejection and its resulting noise rejection as a function of frequency
Figure 11. High-Order DC Noise Rejection

Table 4 summarizes the differences between the three DC noise rejection settings.

 

Table 4. DC Noise Rejection Settings

AC Voltage Measurements

AC signals are typically characterized by rms amplitude, which is a measure of their total energy. RMS stands for root-mean-square; to compute the rms value of a waveform, you must take the square root of the mean value of the square of the signal level. Although most DMMs do this nonlinear signal processing in the analog domain, the FlexDMM architecture uses an onboard DSP to compute the rms value from digitized samples of the AC waveform. The result is quiet, accurate, and fast-settling AC readings. The digital algorithm automatically rejects the DC component of the signal, making it possible to bypass the slow-settling input capacitor. To measure small AC voltages in the presence of large DC offsets, such as ripple on a DC power supply, the FlexDMM architecture offers the standard AC volts mode, in which the coupling capacitor eliminates the offset and the FlexDMM architecture uses the most sensitive range.

The rms algorithm used by the FlexDMM architecture requires only four periods (cycles) of the waveform to obtain a quiet reading. For example, it requires a measurement aperture of 4 ms to accurately measure a 1 kHz sine wave. The advantage brought about by this technique extends to system performance. With traditional DMMs, it is necessary to wait for an analog Trms converter to settle before you make a measurement. With the FlexDMM architecture, there is no Trms converter to settle. The result is faster AC reading rates, and this advantage is realized in systems with switching.

The digital approach to rms computation offers accuracy benefits as well. The algorithm is completely insensitive to crest factor and can deliver exceptionally quiet and stable readings. The FlexDMM architecture guarantees AC accuracy down to 1 percent of full-scale, rather than the 10 percent of full-scale offered by traditional DMMs, and usable readings are obtainable even below 0.1 percent of full-scale.

Current Measurement Architecture

Current measurements often require support for both very small leakage currents and substantially larger load currents. Supporting this wide dynamic range requires careful consideration of protection, signal conditioning, accuracy, and measurement sensitivity. The current measurement architecture used in the NI PXI DMM was designed to address these competing requirements while maintaining reliability and measurement accuracy.

The PXIe-4081 addresses both needs by implementing a novel solid-state current measurement configuration that provides eight DC current ranges from 1 µA to 3 A and six ACRMS current ranges from 100 µA to 3 A. The 1 µA range offers sensitivity down to 1 pA, or 10-12 A. Providing both extremes requires a unique circuit design approach. The challenges of high-voltage or current-overload protection and low-leakage measurement have historically been mutually exclusive. The FlexDMM architecture implements a unique design approach, diagrammed in Figure 12. This greatly simplified figure shows three of the five current ranges used in the PXIe-4081.


Simplified PXIe-4081 Current Signal Conditioning displaying three of the five current ranges used in the PXIe-4081
Figure 12. Simplified PXIe-4081 Current Signal Conditioning

 

Resistance Measurement Architecture

NI PXI DMMs support resistance measurement workflows that address both basic resistance measurements and more demanding scenarios where lead resistance, switching, or offset voltages can introduce error. Two-wire measurements support general-purpose resistance measurements, while 4-wire measurements help remove lead-resistance effects in low-resistance applications.

Offset-Compensated Ohms

Offset-compensated Ohms measurements further improve accuracy in situations where offset voltages are present in the measurement path. By measuring with the current source on and off, the instrument can distinguish the voltage generated by the resistance under test from offset voltages that would otherwise create measurement error. The following applications benefit from offset-compensated resistance measurements:

  • Switching systems using uncompensated reed relays (uncompensated reed relays can have offset voltages greater than 10 µV caused by the Kovar lead material used at the device glass seal)
  • In-circuit resistance measurements (for example, power supply conductors being measured for resistance while the circuit under test has power applied)
  • Measuring the source resistance of batteries, dynamic resistance of forward-biased diodes, and so on

In Case 1, a test system is often built with switching optimized for tasks other than resistance measurements. For example, reed relays are common in RF test systems because of their predictable impedance characteristics and high reliability. In such a system, you may also want to measure resistances of units under test, and the reed relays may already exist in the system.

In Case 2, an example would be measuring the resistance of a power supply bus wire with the power on. (Note: You need to exercise great care when performing these tests.) Assume the resistance is in the range of 10 mΩ. If there is 100 mA flowing through this resistance, the voltage drop is shown in the following formula:

V = 100 mA x 10 m = 1 mV
 

A DMM without offset compensation on the 100 range interprets this as 1 Ω because it thinks this voltage is being generated by its internal 1 mA current source passing through the wire you are measuring. It cannot tell the difference. With the FlexDMM architecture and offset-compensated Ohms enabled, the 1 mV offset is distinguished and rejected, and the correct value of resistance is returned. 

Figure 13. Diagram with the current source on. Figure 14. Diagram with the current source off
Figure 13. First Cycle with Current ON Figure    
         Figure 14. Second Cycle with Current OFF


This measurement involves two cycles. One is measured with the current source on, as shown in Figure 13. The second is with the current source off, as shown in Figure 14. The net result is the difference between the two measurements. Because the offset voltage is present in both cycles, it is subtracted out and does not enter into the resistance calculation, as shown in the following equation.

​VOCO = VM1 - VM2 = (ISRX + Vthermal) – Vthermal= ISRX

therefore:

RX = VOCO/IS

Isolated Digitizer Architecture

On supported models, NI PXI DMMs combine precision DMM functionality with integrated waveform acquisition capability. This feature allows engineers to investigate transients, capture waveform behavior, and perform additional analysis without introducing separate instrumentation into the test system.

Waveform acquisition can accelerate debugging and characterization by allowing engineers to move from scalar measurements to analysis of transients and other non-repetitive, high-voltage waveforms in both the time and frequency domains, while maintaining the same PXI setup and software workflow. The PXIe-4081 represents the highest-performance implementation of this capability within the NI PXI DMM.

For example, an application in the automotive industry is the measurement of the flyback voltage on an ignition coil. The ignition coil, which creates the high voltages used to drive the spark plugs in the engine, is made up of a primary coil and a secondary coil. The secondary coil generally has many more turns of wire than the primary coil because the turns ratio times the voltage applied to the primary coil determines the output voltage. When the current is suddenly commutated off, the collapse of the magnetic field induces a large voltage (+20,000 V) onto the secondary coil. This voltage is then routed to the spark plugs.

Because the voltages are so high on the secondary coil, tests are actually made on the primary coil. The flyback waveform is usually on the order of 10 µs with a peak voltage of 40 to 400 V, depending on the ignition coil. The common measurements made on this waveform are peak firing voltage, dwell time, and burn time. Using the FlexDMM digitizer capability and software analysis, you can build a flyback voltage measurement system using only the PXIe-4081.

Benefits of an Isolated Digitizer

Isolation enables measurement of signals in the presence of common-mode voltages and helps reduce errors caused by unwanted ground paths. On select NI PXI DMMs with an onboard digitizer, isolation is useful when capturing transient behavior in systems where the signal of interest is not referenced to the same ground as the measurement system. Isolation provides the following benefits:

  • Improved rejection—Isolation improves rejection of common-mode voltages that are not part of the signal being measured.
  • Improved safety—Isolation provides a barrier that can help support floating measurements and withstand transient voltage events within the instrument’s rated limits.
  • Improved measurement integrity—Isolation helps reduce ground-loop effects, a common source of noise and error in measurement systems.

Conclusion

NI PXIe-408x digital multimeters are built upon a common architectural foundation that combines precision measurement performance with the flexibility of modular PXI systems. Together, FlexDMM architecture and FlexADC provide the architectural foundation behind these DMMs. While FlexADC serves as the core measurement engine, FlexDMM architecture integrates that capability with precision analog design, calibration technologies, signal processing, and software-defined analysis to create a flexible measurement platform. The result is PXI DMMs that deliver high accuracy, speed, and scalability across applications ranging from benchtop characterization to complex automated test systems.

CharacteristicDual SlopeCB Ramp DownΣ - AFlexADC™
SpeedGoodBetterGoodBest
Speed-Resolution FlexibilityGoodBetterGoodBest
LinearityBetterBestBetterBest
Noise RejectionBetterBetterBetterBest
Cost/PerformanceGoodBetterBetterBest
ConditionTraditional 7½-Digit DMM (1 Year)PXIe-4081 7½-Digit DMM (2 Year)
​Measurement within 18 to 28 °C

160 µV

65 µV

Measurement at 50 °C without Self-Calibration

600 µV

111 µV

​Measurement at 50 °C with Self-Calibration

600 µV (no self-calibration available)

80 µV

Calibration OptionApplicationWhenPerformance Benefit
​“Factory” Calibration​Recalibrate time drift of  onboard references
Corrects for AC flatness drift for all functions

Every 2 years

To full specifications

Self-Calibration7½-digit precision— recalibrates measurement path and ADC for VDC, resistance, diode, and digitizer

90 days or for temperature change >1 °C

To specifications on VDC, resistance, diode, and digitizer functions over FULL operating temperature range

DC Noise Rejection SettingLowest Frequency for Noise RejectionHigh-Frequency Noise Rejection
Normal1/tapertureGood
Second-Order2/tapertureBetter
​High-Order4/tapertureBest >110 dB rejection