The NI-9222 with screw terminal has been evaluated as Ex nA IIC T4 Gc equipment under DEMKO Certificate No. 07 ATEX 0626664X and is IECEx UL 14.0089X certified. The NI-9222 with BNC has been evaluated ...

Channel-to-channel None Channel-to-earth ground Continuous 250 Vrms, Measurement Category II Withstand 3,000 Vrms, verified by a 5 s dielectric withstand test Parent topic: Safety Guidelines

This document explains how to connect to the NI-9222 . Note Before you begin, read the NI-9222 Safety, Environmental, and Regulatory Information document on ni.com/manuals and complete the software and ...

DSUB or spring terminal connectivity 50 Hz / 60 Hz noise rejection 250 V RMS , CAT II , channel-to-earth isolation (spring terminal); 60 V DC , CAT I , channel-to-earth isolation (DSUB) The NI-9216 RTD ...

EX+ VEX+ VEX- V SUP+ V SUP - NI 9218 + - AI+ AI - /EX - The NI 9218 provides optional 12 V sensor excitation. To use the 12 V excitation, connect a 9 VDC to 30 VDC power

EX+ VEX+ VEX- V SUP+ V SUP - NI 9218 NI 9983 AI+ AI- DSUB/LEMO EX- Connecting ±20 mA signals requires the NI 9983 . The NI 9218 provides optional 12 V sensor excitation. To use the 12 V excitation,

NI 9218 NI 9987 AI+ + - AI- DSUB/LEMO Connecting ±60 V signals requires the NI 9987 . Parent topic: Measurement Types Related reference NI 9987 Pinout

EX+ EX- VEX+ VEX- V SUP+ V SUP - NI 9218 + - AI+ AI - You must connect AI- to EX- on the NI 9218 . The NI 9218 provides optional 12 V sensor excitation. To use the 12

3a 10 9 6 7 8 — AI+ AI- — — 3b 4 5 2 1 EX- — — — EX+ 4 5 1 2 3a 3b 6 7 8 9 EX+ — EX- — — — AI+ AI- —

Located at n i . c o m / manuals Installs with the software CompactRIO NI CompactDAQ REL A TED INFORM A TION NI 9221 Datasheet NI-RIO Help LabVIEW FPG A Help NI 9221 Datasheet NI-DAQmx Help LabVIEW Help ...

The following specifications are typical for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to the AI- signal on each channel unless otherwise noted. Caution Do not operate ...

To meet the EMC guidelines in this document and the Electromagnetic Compatibility specifications in the NI-9226 Datasheet , you must use the NI 9940 connector backshell kit. Figure 2. NI 9940 Connector ...

Screw Terminal connectivity 50 Hz / 60 Hz noise rejection 250 V RMS , CAT II , channel-to-channel isolation The NI-9224 is an analog input module for use in CompactDAQ or CompactRIO systems. Each channel ...

Connect only voltages that are within the following limits. Channel-to-channel None Channel-to-earth ground Continuous 60 V DC, Measurement Category I Withstand up to 3,000 m 1,000 V RMS, verified by a ...

This document explains how to connect to the NI-9226 . In this document, the NI-9226 with spring terminal and the NI-9226 with DSUB are referred to inclusively as the NI-9226 . Note Before you begin, read ...

Make sure that devices you connect to the NI-9226 are compatible with the module specifications. Parent topic: Overview

Ensure that the environment in which you are using the NI-9226 meets the following specifications. Operating temperature (IEC 60068-2-1, IEC 60068-2-2) -40 °C to 70 °C Operating humidity (IEC 60068-2-30) ...

This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable ...

The NI-9226 is suitable for use in hazardous locations; , and hazardous locations; and nonhazardous locations only. Follow these guidelines if you are installing the NI-9226 in a potentially explosive ...

Channel-to-channel None Channel-to-earth ground Continuous 250 Vrms, Measurement Category II Withstand 3,000 Vrms, verified by a 5 s dielectric withstand test Parent topic: Safety Guidelines