Some products are approved for marine (shipboard) applications. To verify marine approval certification for a product, visit ni.com/product-certifications , search by model number, and click the appropriate ...

This document explains how to connect to the NI-9216 . In this document, the NI-9216 with spring terminal and the NI-9216 with DSUB are referred to inclusively as the NI-9216 . Note Before you begin, read ...

NC EX0+ RTD0+ RTD0- COM EX1+ RTD1+ RTD1- COM EX2+ RTD2+ RTD2- COM EX3+ RTD3+ RTD3- COM NC NC EX4+ RTD4+ RTD4- COM EX5+ RTD5+ RTD5- COM EX6+ RTD6+ RTD6- COM EX7+ RTD7+ RTD7- COM NC 1 2 3 4

Each channel has an open thermocouple detection (OTD) circuit, which consists of a current source between the TC+ and TC- terminals.

What to Use NI-9213 spring-terminal connector 0.08 mm 2 to 1.0 mm 2 ( 28 AWG to 18 AWG ) copper conductor wire with 7 mm ( 0.28 in. ) of insulation stripped from the end Flathead screwdriver with a

This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable ...

The NI-9213 is suitable for use in Class I, Division 2, Groups A, B, C, D, T4 hazardous locations; Class I, Zone 2 , AEx nA IIC T4 Gc and Ex nA IIC T4 Gc hazardous locations; and nonhazardous locations ...

The NI 9213 has been evaluated as Ex nA IIC T4 Gc equipment under DEMKO 07 ATEX 0626664X and is IECEx UL 14.0089X certified. Each NI 9213 is marked II 3G and is suitable for use in Zone 2 hazardous

If hazardous voltages are connected to the product, take the following precautions. A hazardous voltage is a voltage greater than: 30 V RMS , 42.4 V peak , or 60 V DC in DRY LOCATIONS 16 V RMS , 22.6

If your application is subject to high vibration, NI recommends that you use the NI-9940 backshell kit to protect connections to the NI-9213 . Parent topic: Overview

The following specifications are typical for the range -40 °C to 70 °C unless otherwise noted. The specifications are for the NI 9214 used in conjunction with an TB-9214 . Caution Do not operate the NI-9214 ...

The following specifications are typical for the range -40 °C to 70 °C unless otherwise noted. Accuracy within typical use can vary based on chassis, mounting parameters, other modules present in the system, ...

This document explains how to connect to the NI-9215 . In this document, the NI-9215 with screw terminal , NI-9215 with spring terminal , and NI-9215 with BNC are referred to inclusively as the NI-9215 ...

If you need to clean the module, wipe it with a dry towel. Tip For two-dimensional drawings and three-dimensional models of the C Series module and connectors, visit ni.com/dimensions and search by module ...

AI0+ AI0- AI1+ AI1– AI2+ AI2– AI3+ AI3– NC COM 0 1 2 3 4 5 6 7 8 9 AI0+ AI0– AI1+ AI1– AI2+ AI2– AI3+ AI3– NC COM 0 1 2 3 4 5 6 7 8 9

Ensure that the environment in which you are using the NI 9215 meets the following specifications. Operating temperature (IEC 60068-2-1, IEC 60068-2-2) -40 °C to 70 °C Operating humidity (IEC 60068-2-78) ...

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications ...

What to Use TB-9212 with mini TC Shielded thermocouple Clamp-on ferrite bead (part number 781233-01 ) What to Do Plug the thermocouple into the thermocouple input on the TB-9212 with mini TC . Install ...

Showing 81-100 of 28307 results
5 of 1416 pages